BioTek Synergy MX Multi-Mode Microplate Reader
Detection Modes
Absorbance, Fluorescence Intensity, Luminescence
Microplate Formats
1- to 384- well
Read Types
Endpoint, Kinetic, Spectral Scanning, Well-Area Scanning
- Fully monochromator-based with quadruple grating architecture for ultra fine-tuned performance
- Compatible with Take3 Multi-Volume Plate with 2 µL microspots for low volume assays
- High performance fluorescence (top read only), absorbance and dedicated luminescence detection system accommodates all life-science research microplate applications
- Automated Z-height adjustment accommodates variable microplate heights and liquid levels with 100 μm resolution
- Variable bandpass selection system adds flexibility
- Cuvette port to measure standard 1-cm path cuvettes
- Controlled by the powerful BioTek Gen5 software, facilitating easy data collection, analysis, exporting, and reporting
The absorbance specifications of the Synergy MX are impressive. It utilizes a Xenon flash lamp as its light source and a double monochromator for wavelength selection, covering a range from 230 to 999 nm in 1 nm increments. The device ensures photometric accuracy of < 1% at 2.0 OD and repeatability of < 0.5% at 2.0 OD, with a dynamic range from 0 to 4.0 OD. For fluorescence intensity measurements, the device also uses a Xenon flash lamp and a dedicated red-shifted PMT detector, with a wavelength range of 250 to 900 nm and a remarkable sensitivity level. The luminescence capabilities of the Synergy MX are equally robust, featuring a low noise PMT detection system and a dynamic range of < 6 decades, with exceptional sensitivity for detecting as low as 10 amol of ATP.
This microplate reader can accommodate formats from 1 to 384-well plates and PCR plates, and offers various read methods such as endpoint, kinetic, spectral scanning, and well-area scanning. It also includes plate shaking features and precise temperature control from Ambient +4°C to 65°C. Controlled by BioTek's powerful Gen5 software, the Synergy MX facilitates easy data collection, analysis, exporting, and reporting, making it an invaluable tool for modern laboratories engaged in life-science research. Please Note: This is the SMATCL model of the Synergy MX, which DOES NOT feature bottom fluorescence reading or a dual reagent dispenser.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioTek Synergy MX Multi-Mode Microplate Reader includes:
- Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- ABS, FLD, LUM, and Self Test Calibration Report
- Interface Cable
- Operator's Manual
- Full 180 Day Part and Labor Warranty
BioTek Synergy MX Multi-Mode Microplate Reader Specifications
|
Absorbance Specifications |
|
|
Light Source |
Xenon flash lamp |
|
Wavelength Selection |
Double monochromator |
|
Wavelength Range |
230 to 999 nm, in 1 nm increments |
|
Bandpass |
2 nm (230-285 nm), 4 nm (>285 nm) |
|
Pathlength Correction |
Yes |
|
Monochromator Wavelength Accuracy |
± 2 nm |
|
Monochromator Wavelength Repeatability |
± 0.2 nm |
|
Dynamic Range |
0 to 4.0 OD |
|
Photometric Resolution |
0.0001 OD |
|
Photometric Accuracy |
< 1% at 2.0 OD typical |
|
Photometric Linearity |
< 1% from 0 to 3.0 OD |
|
Photometric Repeatability |
< 0.5% at 2.0 OD |
|
Reading Speed |
96 wells: 11 seconds 384 wells: 22 seconds |
|
Fluorescence Intensity Specifications |
|
|
Light Source |
Xenon flash lamp |
|
Detector |
Dedicated red shifted PMT |
|
Wavelength Selection |
Double grating monochromators (top read only) |
|
Wavelength Range |
250 to 900 nm |
|
Dynamic Range |
5 decades |
|
Sensitivity (Top Read Only) |
Fluorescein 2 pM typical (0.2 fmol/well, 384-well plate) |
|
Luminescence Specifications |
|
|
Dynamic Range |
< 6 decades |
|
Sensitivity |
10 amol ATP typical (flash) |
|
Detection System |
Low noise PMT |
|
Wavelength Range |
300 to 700 nm |
|
General Specifications |
|
|
Microplate Formats |
1- to 384- well plates and PCR plates |
|
Read Methods |
End point, kinetic, spectral scanning, well-area scanning |
|
Plate Shaking |
Yes |
|
Temperature Control |
Ambient +4°C to 65°C, ± 0.5°C at 37°C |
|
Dimensions (W x D x H) |
17 x 20.9 x 15 in. (43.5 x 53.1 x 38.1 cm) |
|
Weight |
70 lb. (31.5 kg.) |
|
Electrical Requirements |
100-240V, 50/60Hz |
|
Computer Specifications |
|
| Processor | Intel Core I5-8350U CPU |
| RAM | 16.0 GB |
| Hard Drive | 512GB Internal SSD |
| Operating System | Windows 11 Professional, 64-bit |
- Verify that the PCB and power supply output proper voltages to all internal components.
- Examine the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal readability.
- Inspect all buttons for responsiveness, confirming full functionality and consistency.
- Assess the plate drawer for proper clearance and smooth operation to ensure accurate plate handling.
- Confirm seamless communication between the plate reader and computer software for data integrity.
- Inspect internal wiring and electrical connections for wear or potential faults, addressing any issues found.
- Examine the lamp, replacing it as needed to maintain proper illumination and performance.
- Execute the required validation protocol to confirm system accuracy and compliance.
- Apply all current updates to the data station and operating system to ensure compatibility and stability.
- Conduct a final cosmetic inspection to verify that all external surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify PCB and power supply output proper voltages | Completed | |
| Check display for brightness, replace dim or dead pixels | Completed | |
| Inspect buttons for full functionality and responsiveness | Completed | |
| Assess plate drawer clearance for accurate plate handling | Completed | |
| Confirm communication between plate reader and software | Completed | |
| Inspect internal wiring and electrical connections for wear | Completed | |
| Examine lamp and replace as necessary for illumination | Completed | |
| Execute validation protocol for system accuracy | Completed | |
| Apply all current updates to data station and OS | Completed | |
| Conduct cosmetic inspection to meet appearance standards | Completed |



Click Here to View the Technical Specs for the BioTek Synergy MX Multi-Mode Microplate Reader