VWR Gravity Convection Incubator, Cat # 414004-620
The VWR Gravity Convection Incubator is for basic incubation with reduced air changes. With a temperature range of an ambient +8C to 65C it is accurate within 0.3C at 37C and a heating time of 30 min. to 37C and 45 min. to 50C. It provides a safe environment while reducing drying samples out. The temperature sensor allows for repeatability and stability over an extended period of time. It has a digital backlit LCD display for easy viewing.
All of our equipment is fully tested to perform at factory specifications before being shipped.
> Shelf
> Power Cord
> Full 180 Day Parts and Labor Warranty
VWR Gravity Convection Incubator, Cat # 414004-620 Specifications
|
Load per Rack |
15 kg (33 lbs.) |
|
Permitted Total Load |
25 kg (55 lbs.) |
|
Wall Clearance |
Rear: 15 cm (5.9) Side: 15 cm (5.9") |
|
Exhaust Duct (Outer) |
4 cm (1.6") |
|
Temperature Range |
Ambient Temperature +8C to 65C |
|
Temperature Accuracy |
within 0.3C at 37C |
|
Temperature setting sensitivity |
0.1C |
|
Temperature variation |
0.7C at 37C, 1.2C at 50C |
|
Heating Time |
30 min. to 37C, 45 min. to 50C |
|
Temperature Recovery (Door Open 30 Sec.) |
11 min. to 37C 16 min. to 50C |
|
Volume |
50 L (1.75 cu. Ft) |
|
Interior Dimensions (L x W x H) |
54.5 54.1 70.1 cm (21 7/16 21 5/16 27 5/8 in) |
|
Electrical |
120V, 60Hz, 200W |
|
Shipping weight |
53 kg (118 lbs.) |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and CO₂ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |