VWR Digital Multi Heat Block - 12621-096
Temperature Range
Ambient +5C to 120C
Temperature Stability
± 0.1°C at 37°C
Temperature Uniformity
± 0.1°C at 37°C
- Accommodates 4 Blocks for versatile sample heating
- High-Wattage Heating Elements ensure efficient performance
- Interchangeable Blocks support various tube sizes from 0.25mL to 50mL
- Thermometer Wells in Each Block for precise temperature monitoring
- Compact and Durable Design optimized for laboratory use
- Reliable Digital Control for accurate operation
All of our equipment is fully tested before being shipped.
The VWR Digital Multi Heat Block - 4 includes:
- > Operator's Guide
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
- > Please Note: Heat blocks are NOT INCLUDED with this Dry Bath Heater. If you require heat blocks, please inquire
VWR Digital Multi Heat Block Specifications
|
Temperature Range |
Ambient +5C to 120C |
|
Temperature Stability |
0.1C |
|
Temperature Uniformity |
0.1C |
|
Dimensions (L x W x H) |
16.9 x 8 x 3.5 in (42.9 x 20.3 x 8.9 cm) |
|
Electrical Requirements |
120V, 50/60Hz, 2.6A, 310W |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |



Click Here to View the Technical Specs for the VWR Multi Heat Block