VWR Analog Heat Block, Cat. No. 13259-030
VWR Analog Heat Block 100 Watt, Multi-purpose unit. Adjustable high and low thermostats provide reproducible and accurate temperature control with a stability of 2C. Temperature range is 5C above ambient to 150C. Please Note: A heat block is NOT INCLUDED with this Dry Bath Heater. If you require a heat block, please inquire.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR Analog Heat Block, Cat. No. 13259-030 includes:
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
- > Please Note: A heat block is NOT INCLUDED with this Dry Bath Heater. If you require a heat block, please inquire
VWR Analog Heat Block, Cat # 13259-030 Specifications
|
Temperature Range |
5C above ambient to 150C |
|
Temperature Stability |
2C |
|
Heat Up Time to 100C |
45 minutes |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |
Click Here to View all the Technical Specs for the VWR Heat Block


Click Here to View the Technical Specs for the VWR Heat Block