VWR 76019-132 Centrifuge w/ 15ml Rotor
Maximum Speed
4,500 RPM
Maximum RCF
2,490 x g
Maximum Capacity
8 x 15 mL
- Digital control system provides straightforward adjustment and monitoring of centrifugation parameters
- Digital display allows users to view RPM or RCF along with run time
- RPM/RCF conversion button enables convenient switching between speed measurements
- Pulse operation supports quick-spin applications
- Brushless DC motor drive provides reliable operation with reduced maintenance requirements
- Adjustable timer supports timed runs from 30 seconds to 99 minutes or continuous operation
- Secure lid lock helps prevent access during centrifugation
- Over-speed detection provides additional operational protection
- Automatic internal diagnostics help monitor proper centrifuge operation
- Compact benchtop design conserves valuable laboratory workspace
The VWR 76019-132 Centrifuge with 15 mL Rotor is a compact, general-purpose centrifuge designed for routine research and clinical laboratory applications. Equipped with an 8 x 15 mL fixed-angle rotor, the centrifuge provides variable speeds from 300 to 4,500 RPM and is powered by a brushless DC motor for reliable operation. A digital display allows users to monitor either RPM or RCF along with run time, while a dedicated RPM/RCF conversion button simplifies switching between operating parameters.
Designed for convenient and safe everyday centrifugation, the 76019-132 features pulse operation for quick spins, a secure door interlock, over-speed detection, and automatic internal diagnostics. Its compact footprint makes it well suited for laboratories requiring dependable general-purpose centrifugation without occupying excessive bench space.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR 76019-132 Centrifuge w/ 15ml Rotor includes:
- 8 x 15 ml Rotor
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
VWR 76019-132 Centrifuge w/ 15ml Rotor Specifications
| Maximum Speed | 4,500 RPM |
| Maximum RCF | 2,490 x g |
| Maximum Capacity | 8 x 15 mL |
| Temperature Range | N/A |
| Timer | 30 seconds to 99 minutes, with continuous run |
| Display | Digital |
| Dimensions (W x D x H) | 11 x 14.3 x 10.4 in. (27.9 x 36.3 x 26.4 cm) |
| Weight | 13.2 lb. (6 kg) |
| Electrical Requirements | 100-240V, 50/60Hz, 70W |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |




Click Here to View the Technical Specs for the VWR 76019-132 Centrifuge w/ 15ml Rotor