VWR 1300U Gravity Convection Utility Oven Specifications
|
Temperature Range
|
40 to 200C
|
|
Temperature Uniformity
|
10C
|
|
Volume
|
28.3 L (1 cu. ft.)
|
|
Exterior Dimensions (W x D x H)
|
16 x 17 x 18.5 in
|
|
Interior Dimensions (W x D x H)
|
12 x 12 x 12 in
|
|
Electrical
|
120V, 60Hz |
- Verify power supply and PCB outputs are within specified voltage ranges
- Inspect display for any dim or dead pixels, replacing as needed to ensure visibility
- Confirm full functionality and responsiveness of all control buttons
- Examine all internal wiring for frays, cuts, or damage, ensuring secure and reliable connections
- Inspect door seals and gaskets for integrity, replacing any compromised components
- Evaluate door handle and locking mechanism for wear or issues, replacing if performance is impaired
- Inspect and clean internal fan; replace if performance does not meet requirements
- Calibrate set temperature using a NIST-traceable probe for precise accuracy
- Confirm functionality of over-temperature alarm and indicator light
- Verify internal outlet functionality, if applicable, to ensure it meets operational standards
- Inspect internal lighting for full functionality and brightness, if applicable
- Perform a cosmetic inspection to confirm the oven’s exterior and interior meet quality standards
Technician Inspection and Recertification Checklist: Ovens
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Verify power supply and PCB outputs are within specified voltage ranges | Completed | |
| Inspect display for dim or dead pixels; replace if needed | Completed | |
| Confirm full functionality of all control buttons | Completed | |
| Examine internal wiring for damage; ensure secure connections | Completed | |
| Inspect door seals and gaskets, replacing any compromised components | Completed | |
| Evaluate door handle and locking mechanism, replacing if worn | Completed | |
| Inspect and clean internal fan; replace if performance is insufficient | Completed | |
| Calibrate set temperature using a NIST-traceable probe | Completed | |
| Confirm over-temperature alarm and indicator light functionality | Completed | |
| Verify internal outlet functionality, if applicable | Completed | |
| Inspect internal lighting for full functionality and brightness | Completed | |
| Perform cosmetic inspection to confirm quality standards of exterior and interior | Completed |