Thermo Sorvall T1 Benchtop Centrifuge
The Thermo Sorvall T1 Benchtop Centrifuge with the T-20 Microplate Rotor is provides unsurpassed versatility and exceptional performance in a quiet, compact, 1.2 Liter capacity benchtop centrifuge. The Thermo Sorvall T1 delivers consistent separations quickly, with maximum convenience and comfort. This centrifuge features the Auto-Lock rotor exchange system allowing for the user to swap rotors within just seconds. The Thermo Sorvall T1 also features a user-friendly control panel with one-touch access to the most frequently used programs. A brushless induction motor also provides years of service with minimal maintenance. This centrifuge is capable of reaching a maximum speed of 14,600 RPM and a maximum RCF of 23,113 x g. The included T-20 Microplate Rotor reaches a maximum speed of 4,100 RPM and maximum RCF of 2,160 x g.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Sorvall T1 Benchtop Centrifuge includes:
- > T-20 Microplate Rotor & Buckets
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Sorvall T1 Benchtop Centrifuge Specifications
|
Maximum Speed
|
14,600 RPM (4,100 RPM with included rotor) |
|
Maximum RCF
|
23,113 x g (2,160 x g with included rotor) |
|
Maximum Capacity
|
4 x 280 ml or 6 x 100 ml (6 x Microplates with included rotor) |
|
Temperature Range
|
N/A
|
|
Timer
|
30 sec to 99 min; continuous
|
| Acceleration Rates | 5 |
| Braking Rates | 5 + no brake |
|
Dimensions (H x W x D)
|
14.6 x 15.7 x 19.7 in (37 x 40 x 50 cm) |
|
Weight
|
88.1 lb. (40 kg.) |
|
Electrical Requirements
|
120V, 60Hz, 9.5A |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Thermo Sorvall T1 Benchtop Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly





Click Here to View the Technical Specs for the Thermo Sorvall T1 Centrifuge