Thermo Scientific MaxQ 4450 Digital Orbital Incubator Shaker w/ Cooling Coil
Speed Range
15 to 500 RPM
Speed Control Accuracy
± 1 RPM
Temperature Range
Ambient +5°C to 80°C (Use of external chiller required to reach temperatures below ambient)
- Adjustable parameters via membrane keypad with fine incremental control
- Non-volatile memory retains settings during power interruptions and automatically resumes operation
- Flashing display alerts users to power loss events for added process awareness
- Integrated cooling coil enables below-ambient operation when paired with an external chiller
- Drive interrupt safety feature stops shaking when the lid is opened
- Independent over-temperature protection with visual alarm for enhanced sample safety
- Audible and visual alarms notify users of deviations or unbalanced operating conditions
- Automatic shutdown of shaking function during fault conditions until corrected
- Solid-state brushless DC motor ensures reliable, low-maintenance operation
- Audible alarm muting function allows temporary silencing without disabling alerts
- Timer supports both continuous operation and extended timed runs for flexible workflows

The Thermo Scientific MaxQ 4450 Digital Orbital Incubator Shaker w/ Cooling Coil combines precise environmental control with reliable orbital shaking in a compact benchtop design, making it ideal for applications such as cell culture, protein expression, extraction procedures, and solubility studies. Designed for versatility, this model supports both incubated operation and below-ambient temperature workflows when paired with an external chiller.
A PID microprocessor controller allows users to accurately regulate speed, temperature, and time through an intuitive keypad interface, with adjustments available in fine 1 RPM increments. The digital display provides clear system feedback and alerts users to power interruptions, while non-volatile memory preserves setpoints and automatically resumes operation once power is restored.
For enhanced performance, the integrated cooling coil enables operation below ambient conditions when connected to a compatible chiller, expanding the range of applications supported within a single unit. Safety features include a lid-open drive interrupt that immediately stops shaking and an independent over-temperature protection system with visual alerts to safeguard samples.
Built for dependable, everyday use, the MaxQ 4450 offers straightforward operation, precise control, and flexible temperature capabilities, making it a practical solution for laboratories requiring consistent incubation and shaking performance.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific MaxQ 4450 Digital Orbital Incubator Shaker w/ Cooling Coil includes:
- Power Cord
- Operator's Manual
- 180 Day Full Parts and Labor Warranty
- **Please Note: Unit requires connection to chiller for use of cooling functionality.**
Thermo Scientific MaxQ 4450 Digital Orbital Incubator Shaker w/ Cooling Coil Specifications
| Speed Range | 15 to 500 RPM |
| Speed Control Accuracy | ± 1 RPM |
| Temperature Range | Ambient +5°C to 80°C (Use of external chiller required to reach temperatures below ambient) |
| Temperature Uniformity | ± 0.1°C at 37°C |
| Timer | Continuous or timed operation from 0.1 hour up to 999 hours or 0.1 minute to 999 minutes |
| Orbit | 1.9 cm (0.75 in.) |
| Load Capacity | 35 lb. (15.9 kg.) |
| Platform Size | 11 x 13 in. (28 x 33 cm) |
| Overall Dimensions (L x W x H) | 27.16 x 14.09 x 15.75 in. (69 x 35.8 x 40 cm) |
| Weight | 77 lb. (35 kg.) |
| Electrical Requirements | 120V, 50/60Hz, 4.5A |
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |
Downloads



Click Here to View the Technical Specs for the Thermo Scientific MaxQ 4450 Digital Orbital Incubator Shaker w/ Cooling Coil