Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer
Maximum Temperature
540C
Stirrer Speed
60 to 1,200 RPM
Platform Size
7 x 7 in (17.7 x 17.7 cm)
- Durable ceramic top plate offers excellent chemical resistance
- Maintains stable performance for heating and stirring applications
- Ideal for sample preparation, reagent heating, and solvent evaporation
- Compact, low-profile design maximizes bench space efficiency
- Accurate surface temperature control via internal sensor
- Reliable construction ensures long-term laboratory use
Equipped with a bright digital display, this unit continuously monitors actual temperature for accurate results, while its simple interface allows easy adjustment of heating and stirring parameters. The Cimarec SP131325Q is ideal for general-purpose heating, sample preparation, reagent mixing, solvent evaporation, and media preparation. Its compact, low-profile design ensures stability and efficient bench space use, while the ceramic surface and robust construction make it dependable for daily laboratory operation.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer includes:
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer Specifications
| Maximum Temperature | 540C |
| Stirrer Speed | 60 to 1,200 RPM |
| Max. Stirring Volume | 4L |
| Platform Size | 7 x 7 in (17.7 x 17.7 cm) |
| Plate Material | Ceramic |
| Weight | 11.0 lb. (5.0 kg) |
| Dimensions (W x D x H) | 8.2 x 13 x 3.8 in. (20.8 x 33 x 9.7 cm) |
| Electrical Requirements | 120V, 60Hz, 8.9A |
- Inspect Display and Replace Any Dim/Dead Pixels (If Applicable).
- Inspect Ceramic Top for Any Cracks and Replace If Necessary.
- Inspect Functionality of All Buttons/Knobs.
- Use Magnetic Stir Bar Alongside NIST Tachometer to Confirm RPM of Internal Motor (If Applicable).
- Inspect Internal Wiring Inside Case.
- Clean Internal Parts and Vent Any Internal Debris.
- Check Feet to Confirm Plate Sits Properly Balanced.
- Replace All Parts That Were Removed on Internal Inspections.
- Confirm Temperature with NIST Traceable Temperature Probe (If Applicable).
- Ensure Cosmetics Are Acceptable.
Technician Inspection and Recertification Checklist: Hot Plate Stirrers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect Display and Replace Any Dim/Dead Pixels (If Applicable) | Completed | |
| Inspect Ceramic Top for Any Cracks and Replace If Necessary | Completed | |
| Use Magnetic Stir Bar Alongside NIST Tachometer to Confirm RPM of Internal Motor (If Applicable) | Completed | |
| Inspect Internal Wiring Inside Case | Completed | |
| Clean Internal Parts and Vent Any Internal Debris | Completed | |
| Check Feet to Confirm Plate Sits Properly Balanced | Completed | |
| Replace All Parts That Were Removed on Internal Inspections | Completed | |
| Confirm Temperature with NIST Traceable Temperature Probe (If Applicable) | Completed |


Click Here to View the Technical Specs for the Thermo Scientific Cimarec SP131325Q Digital Hot Plate Stirrer