Thermo Scientific A10 Circulating Chiller
The Thermo Scientific A10 Circulating Chiller w/ SC150
Digital Controller provides excellent mixing and temperature stability even as
fluids become viscous. Units consist of a non-CFC air-cooled refrigeration
system, circulation pump, seamless stainless steel bath, work area cover, and a
SC150 digital temperature controller. The A10 Chiller has a temperature range
from -10 to +100C and provides efficient cooling for both open and closed
applications. This chiller features energy-saving power consumption limits, small
footprint and low noise refrigeration circulators, external circulation, and front
drainage port.
The Thermo Scientific A10 is recommended for calibration,
bioreactors, rotary evaporators, condensers, and sample/material testing and
preparation.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific A10 Circulating Chiller includes:
- > Recirculating Chiller Calibration and Preventative Maintenance Report
- > Operator's Manual
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific A10 Circulating Chiller Specifications
|
Temperature Range |
-10C to +100C |
|
Cooling Capacity |
240W |
|
Reservoir Volume |
4 to 6L (1 to 1.5 gal) |
|
Pump Type |
Positive Displacement |
|
Flow Rate |
17L/min (4.5 gpm) |
|
Electrical Requirements |
115V, 60Hz, 11.5A |
| Temperature Stability | 0.02C |
| Maximum Pump Pressure | 4.4 psi (200 mbar) |
| Weight | 60.6 lbs (27.5 kgs) |
- Inspect display for dimming or dead pixels and replace as necessary to maintain clear visual feedback
- Verify functionality of all buttons and input devices, ensuring reliable and responsive control
- Check that all external hosing connections are secure and properly installed to prevent leakage
- Test unit flow rate on a closed-loop system to confirm consistent fluid dynamics
- Examine all external surfaces for leaks, confirming a sealed and contained system
- Inspect internal components for any signs of leaks to ensure system integrity
- Verify power supply output for accurate voltage delivery per operational specifications
- Inspect all internal wiring, checking for any frays, tears, or compromised insulation
- Replace internal filters to maintain optimal fluid purity and system performance
- Confirm that the fan operates at proper speeds to provide effective cooling
- Verify compressor functionality for consistent performance and cooling efficiency
- Test cooling capacity, confirming it meets performance specifications at 20°C
- Measure water pump output to confirm correct PSI for stable fluid flow
- Determine the lowest achievable temperature, verifying against manufacturer standards
- Ensure cosmetic condition meets standards with no damage affecting functionality or aesthetics
Technician Inspection and Recertification Checklist: Recirculating Chiller
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for dimming or dead pixels, replace if necessary | Completed | |
| Verify functionality of all buttons and input devices | Completed | |
| Ensure external hosing connections are secure | Completed | |
| Test flow rate on closed-loop system | Completed | |
| Inspect for external leaks | Completed | |
| Inspect for internal leaks | Completed | |
| Verify power supply output voltage | Completed | |
| Inspect internal wiring for frays or tears | Completed | |
| Replace internal filters | Completed | |
| Confirm fan functionality | Completed | |
| Verify compressor functionality | Completed | |
| Test cooling capacity at 20°C | Completed | |
| Confirm water pump PSI output | Completed | |
| Determine lowest achievable temperature | Completed | |
| Ensure cosmetic condition meets standards | Completed |
Tests Performed on the Thermo Scientific A10 Circulating Chiller
- Cooling capacity test is performed at 20C to confirm chiller functionality
- Flow rate is tested to verify pump is meeting manufacturer specifications
- Chiller is temperature cycled numerous times to ensure longevity of instrument
- Low temperature test is performed to ensure chiller is functioning within temperature range
Preventative Maintenance Performed on the Thermo Scientific A10 Circulating Chiller
- Pump is Inspected for Faults and Signs of Disrepair
- Motor is Inspected for Shorts and Corroded Wiring
- Plumbing is Inspected for Leaks and Corrosion
- Filters are Cleaned and Replaced if Needed
- Tank is Inspected for Leaks and Cleaned
- Internal Circuitry is Inspected for Faulty/Shorted Components and Replaced Accordingly
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Pump is Lubricated as Needed
Thermo Scientific A10 Chiller Manual
Click Here to View all the Technical Specs for the A10