Ohaus FC5706P Frontier Centrifuge
Speed Range
200 to 6,000 RPM
Maximum RCF
4,427 x g
Maximum Capacity
6 x 50 ml
- Compact, multi-function centrifuge for general lab use
- Backlit double-line LCD for easy viewing of parameters
- 99 memory slots for quick access to stored settings
- Mechanical lid lock and imbalance detection for safety
- 10 acceleration and deceleration settings for process control
- Quiet, brushless DC motor for smooth operation
- Small footprint conserves valuable bench space
- CE and FCC certified for regulated lab environments
- Touch-wheel interface allows operation even with gloves
The Ohaus FC5706P Frontier Multi Centrifuge (Recertified) offers dependable performance in a space-saving design, making it well suited for routine laboratory workflows. Operating from 200 to 6,000 RPM and delivering up to 4,427 × g, this unit supports a standard configuration of 6 × 50 mL tubes, with additional rotor options available to accommodate a range of sample formats.
A clear, backlit LCD screen displays both target and real-time run parameters, while onboard memory for up to 99 methods simplifies repeat protocols and improves day-to-day efficiency. The system incorporates a mechanical lid lock for operator safety and manual rotor selection to support multiple applications.
Designed for ease of use, the FC5706P includes adjustable acceleration and braking profiles with 10 selectable settings, allowing greater control over sample handling. Its brushless motor ensures reliable, low-noise operation, and the compact footprint makes it a practical addition to labs where bench space is limited. Backed by CE and FCC certifications, this recertified unit delivers a cost-effective solution without compromising performance.
The Ohaus FC5706P Frontier Centrifuge includes:
- Power Cord
- Operator's Manual
- 180 Day Full Parts and Labor Warranty
- **Please Note: Rotor sold separately, please inquire with our customer service team.**
Ohaus FC5706P Frontier Centrifuge Specifications
| Speed Range | 200 to 6,000 RPM |
| Maximum RCF | 4,427 x g |
| Maximum Capacity | 6 x 50 ml |
| Program Memories | 99 |
| Acceleration Gear | 10 |
| Deceleration Gear | 10 |
| Display | LCD with Backlight |
| Lid lock | Mechanical |
| Noise Level | ≤ 56 ± 2 dB(A) |
| Power | AC Power (Included) |
| Rotor Recognition Method | Manual |
| Running Time | Continuous; 10 sec to 99 h 59 min 59 sec; Quickspin |
| Safety Certification | CE; FCC |
| Working Environment | 41°F - 95°F, 80%RH, non-condensing (5°C – 35°C, 80%RH, non-condensing) |
| Weight | 31.3 lb (14.2 kg) |
| Dimensions (L x H x W) | 18.3 x 10.7 x 12.7 in (46.5 x 27.1 x 37.0 cm) |
| Electrical Requirements | 100 - 240V, 50/60Hz ,70W |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Downloads