New Brunswick Innova 40 Incubator Shaker
Speed Range
25 to 500 RPM (Depending on Load Weight)
Temperature Range
5°C Above Ambient to 80°C
Orbit Size
0.75 in. (1.9 cm)
- Triple-eccentric counterbalanced drive provides smooth, dependable shaking performance
- Solid-state brushless motor supports reliable operation with reduced maintenance
- Programmable controller supports continuous, timed, and multi-step operation
- ±1 RPM speed control accuracy provides precise shaking performance
- ±0.1°C temperature control accuracy at 37°C supports consistent incubation conditions
- Large backlit LCD provides clear monitoring of operating parameters
- Transparent insulated lid allows convenient viewing of samples during operation
- Drive interrupt system stops shaking when the lid is opened
- Controlled acceleration and deceleration help minimize splashing and mechanical stress
- Imbalance detection automatically stops the motor when an unbalanced condition is detected
- Visual and audible alarms alert users to operating deviations and completed timed runs
- RS-232 connectivity supports remote control, monitoring, and data logging
- Compact benchtop design helps conserve laboratory workspace
- Accommodates flasks up to 3 L for flexibility across different culture volumes

The New Brunswick Innova 40 Incubator Shaker is a benchtop orbital shaker designed for dependable temperature-controlled shaking in research and laboratory applications. Its heavy-duty triple-eccentric counterbalanced drive provides smooth orbital motion and reliable operation, while accommodating flasks up to 3 L. The Innova 40 provides temperature control from 5°C above ambient to 80°C and is available with either a 0.75 in. (1.9 cm) orbit. A transparent, insulated lid allows samples to remain visible during operation without occupying additional bench space.
The programmable controller supports continuous operation, timed runs, and multi-step programs incorporating changes in speed and temperature. Integrated safeguards include a drive interrupt when the lid is opened, controlled acceleration and deceleration to minimize splashing, imbalance detection, and independent over-temperature protection. Visual and audible alarms provide notification of speed or temperature deviations, power failure, an open lid, and completion of timed runs, while an RS-232 interface supports connection to a computer for additional control.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The New Brunswick Innova 40 Incubator Shaker includes:
- 18 x 18 in. Platform
- NIST Calibration and Preventative Maintenance Report
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
New Brunswick Innova 40 Incubator Shaker Specifications
| Speed Range | 25 to 500 RPM (Depending on Load Weight) |
| Speed Control Accuracy | ±1 RPM |
| Temperature Range | 5°C Above Ambient to 80°C |
| Temperature Control Accuracy | ±0.1°C at 37°C |
| Timer | Timed or Continuous Operation |
| Program Modes | Continuous, Timed, Multi-Step Programming, and RS-232 Control |
| Orbit Size | 0.75 in. (1.9 cm) |
| Platform Size | 18 x 18 in. (46 x 46 cm) |
| Chamber Dimensions (W x D x H) | 20.3 x 21.5 x 14 in. (51.4 x 54.4 x 35.6 cm) |
| Overall Dimensions (W x D x H) | 22 x 30 x 24 in. (55.9 x 76.2 x 61 cm) |
| Weight | 133 lb. (60 kg) |
| Electrical Requirements | 120V, 60Hz, 800VA |
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |
Downloads