New Brunswick C1 Orbital Shaker
Speed Range
50 to 400 RPM
Orbit Range
0.75 in.
Control
Analog
- Analog control for simple and reliable speed adjustments
- 11 x 13-inch platform accommodates various vessel sizes
- Designed for operation in temperatures from 5°C to 45°C
- Compact and durable construction for long-term use
- Suitable for general laboratory shaking applications
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The New Brunswick C1 Orbital Shaker includes:
- > 11" x 13" Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
New Brunswick C1 Orbital Shaker Specifications
|
Orbit Range |
0.75 in. |
|
Speed range |
50 to 400 RPM |
|
Control |
Analog |
|
Platform Size |
11 W x 13 D in. |
|
Operating conditions |
5C to 45C |
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Validate stable, correct voltage outputs from power supply and PCB | Completed | |
| Inspect display for dim, flickering, or dead pixels; replace as needed | Completed | |
| Test button functionality for responsiveness and accuracy across all settings | Completed | |
| Inspect bearings and housing for wear; replace as needed to prevent disruption | Completed | |
| Examine drive belt for cracking, fraying, or material breakdown; replace as required | Completed | |
| Measure and adjust motor drive belt tension to optimal specifications | Completed | |
| Verify onboard timer accuracy; replace if it deviates from expected performance | Completed | |
| Calibrate shaker rotation speed using a NIST-traceable tachometer | Completed | |
| Run 8-12 hour continuous test; observe platform for smooth, consistent motion | Completed | |
| Perform comprehensive cosmetic inspection for quality standards compliance | Completed |