Mettler Toledo PG802-S Balance
The Mettler Toledo PG802-S Balance is a highly durable and precise digital lab scale, ideal for a wide range of weighing needs. It boasts a maximum capacity of 810g with a readability of 0.01g and a repeatability of 0.008g, ensuring accurate measurements. The balance is equipped with an internal calibration system, the FACT (Fully Automatic Calibration Technology), allowing for easy and automatic calibration with an internal weight. This rugged balance is chemically resistant and features a user-friendly keypad designed for one-handed operation. Its large digital display and DeltaTrac graphic indicator provide clear visibility and a visual representation of the remaining capacity.
The PG802-S offers a variety of built-in functions, including piece counting, percent weighing, formula weighing, and dynamic weighing, enhancing its versatility. It also supports below-the-balance weighing with a hook, adding to its functionality. With its RS232C interface, this balance ensures seamless integration and efficient data collection. The balance has a quick stabilization time of 1 to 2 seconds and exhibits minimal temperature drift sensitivity (± 6 ppm/°C) and long-term stability (± 0.005%). Integrated short-form operating instructions are included to aid daily operations. The Mettler Toledo PG802-S Balance is a reliable and robust choice for laboratories requiring precise and versatile weighing solutions.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Mettler Toledo PG802-S Balance includes:
- Calibration Label
- Power Cord
- Full 180 Day Parts and Labor Warranty
Mettler Toledo PG802-S Balance Specifications
| Maximum Capacity | 810g |
| Readability | 0.01g |
| Repeatability | 0.008g |
| Linearity | ± 0.01 g |
| Calibration Type | Internal |
| Stabilization Time | 1 to 2 seconds |
| Temperature Drift Sensitivity | ± 6 ppm/°C |
| Long-Term Stability | ± 0.005% |
| Weighing Pan Dimensions | 6.5 x 6.5 in. (16.5 x 16.5 cm) |
| Overall Dimensions (W x D x H) | 8.9 x 14.2 x 3.7 in. (22.6 x 36 x 9.5 cm) |
| Weight | 12.1 lb. (5.5 kg) |
| Electrical Requirements | 9.5-17.5V, 50/60Hz, 10VA Power Supply |
- Verify power supply output for precise voltage delivery in accordance with equipment specifications
- Assess full functionality of display and control buttons, replacing any defective or deteriorating components as necessary
- Confirm balance is properly leveled to ensure measurement accuracy and stability
- Examine windscreen glass for cracks or chips, replacing if structural integrity is compromised
- Inspect all internal wiring and electrical connections for secure attachment and check for any loose connections or damaged connectors
- Check power cord integrity, ensuring there are no frays, cuts, or damage affecting function or safety
- Execute internal or external calibration to confirm accuracy of weighing measurements
- Review cosmetic condition, confirming no visible damage or wear that affects functionality
- Verify successful taring of the weighing pan to zero out prior measurements accurately
- Ensure RS232/printer communication functions correctly, validating connection and data transmission (if applicable)
Technician Inspection and Recertification Checklist: Balances
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify power supply output voltage | Completed | |
| Assess display and control button functionality, replacing faulty components | Completed | |
| Confirm balance leveling for measurement accuracy | Completed | |
| Inspect windscreen glass for cracks, replace if necessary | Completed | |
| Check internal wiring and electrical connections for security | Completed | |
| Inspect power cord for frays or damage | Completed | |
| Execute internal/external calibration for measurement accuracy | Completed | |
| Review cosmetic condition for visible damage | Completed | |
| Verify weighing pan tares successfully | Completed | |
| Ensure RS232/printer communication is functional (if applicable) | Completed |