Labline 3527 Orbital Incubator Shaker
Speed Range
40 to 400 RPM
Temperature Range
Ambient to 65°C
Temperature Uniformity
± 0.6°C
- Combines a controlled temperature incubator with an orbital shaking mechanism in a compact, benchtop design
- Ensures consistently even heating through constant air circulation via a speed blower
- Digital display for accurate monitoring of shaker speed
- Operates on a 3/4" circular motion utilizing nine isolated bearings to enhance durability and safety
- Dual mode operation allows for both continuous shaking or timed shaking up to 60 minutes
- Chamber dimensions of 21" W x 20 3/4" D x 13 3/4" H provide ample space for various sample sizes
- Temperature control accuracy of ± 0.2°C ensures precision in sample conditions
The LabLine 3527 Orbital Incubator Shaker combines a
controlled temperature incubator with an orbital shaking mechanism in a
benchtop design. Expect consistently even heating with the Labline 3527 from constant
air circulation through the chamber's speed blower. Temperature can be
controlled from slightly above ambient to 65°C. A solid state speed controller
allows for variable speeds between 40 and 400 orbits per minute with a digital display
of shaker speed. This shaker operates on a 3/4" circular motion, rotating on nine
isolated bearings to prevent damage to the mechanism in case of accident. The
Labline 3527 can operate continuously or for a timed shaking of up to 60
minutes.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Labline 3527 Orbital Incubator Shaker includes:
- Platform
- NIST Calibration and Preventative Maintenance Report
- Power Cord
- Operator's Guide
- Full 180 Day Parts and Labor Warranty
Labline 3527 Orbital Shaker Specifications
|
Shaker
Motion
|
3/4"
circular motion
|
|
Shaking
Speed
|
40 - 400 rpm
|
|
Temperature
Control
|
± 0.2°C (at
37°C in flask)
|
|
Temperature
Uniformity
|
± 0.6°C (at
37C in flask)
|
|
Timer
|
1 - 60
minutes or continuous
|
|
Chamber
Dimensions
|
21" W x
20 3/4" D x 13 3/4" H
|
|
Overall
Dimensions
|
22
3/16" W x 29 3/4" D x 22 5/8"
|
|
Power
Requirements
|
120 VAC, 50
/ 60 Hz, 600 Watts
|
|
Net Weight
|
135 lbs
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Examine chamber glass for cracks or chips, replace as needed | Completed | |
| Inspect door seals for wear, replace to ensure tight seal | Completed | |
| Check display brightness, replace dim or dead pixels | Completed | |
| Test all buttons for full functionality and responsiveness | Completed | |
| Replace platform assembly bearings for reliable motion | Completed | |
| Inspect drive belt for wear, replace if torn or cracked | Completed | |
| Examine wiring for frays or breakages, repair as needed | Completed | |
| Verify PCB test points, replace faulty components | Completed | |
| Check drive belt tension for proper alignment | Completed | |
| Confirm compressor functionality, clean filter and fins | Completed | |
| Calibrate internal temperature with NIST probe | Completed | |
| Calibrate platform RPM with NIST-traceable tachometer | Completed | |
| Conduct cosmetic inspection for appearance standards | Completed |