Hettich EBA 200 Centrifuge
The Hettich EBA 200 Centrifuge is a compact clinical centrifuge offering performance and safety for small sample volumes. The EBA 200 features a built-in, versatile, 8-place fixed angle rotor for 15 ml tubes, along with an adjustable LCD display for convenient and quick operation. The user may also utilize 3 adjustable messages to notify of completion of the centrifugation run, whether it be audio, visual, or a combination of both.
This rotor reaches a maximum speed of 6,000 RPM.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Hettich EBA 200 Centrifuge includes:
- Hettich 8-Place 15ml Fixed Angle Rotor
- NIST Centrifuge Calibration and Preventative Maintenance Report
- Power Cord
- Full 180 Day Parts and Labor Warranty
Hettich EBA 200 Centrifuge Specifications
|
Maximum Speed
|
6,000 RPM (6,000 with included rotor)
|
|
Maximum RCF |
3,461 x g (3,461 x g with included rotor)
|
|
Maximum Capacity
|
8 x 15 ml (8 x 15 ml with included rotor) |
|
Temperature
Range
|
N/A |
|
Timer
|
1 sec – 99
min, continuous run,
short cycle
mode (impulse button)
|
|
Dimensions (W
x D x H)
|
10.3 x 13.9 x 8.9 in. (26.1 x 35.3 x 22.8 cm)
|
|
Weight
|
19.8 lb. (9 kg)
|
|
Electrical
Requirements
|
100-127V, 50/60Hz, 1.0A |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Hettich EBA 200 Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly




Click Here to View the Technical Specs for the Hettich EBA 200 Centrifuge