Fisherbrand GT05LFEEFSA Undercounter Lab Freezer
Capacity
4.5 cu. ft. (127 L)
Temperature Range
-20 to -10°C (-4 to 14°F)
Defrost Type
Manual
- Compact undercounter design helps maximize available laboratory space
- 4.5 cu. ft. storage capacity accommodates a variety of laboratory materials
- Controlled low-temperature storage supports temperature-sensitive samples
- Manual defrost system provides straightforward operation and maintenance
- Lockable door helps secure stored samples and materials
- Space-efficient footprint is well suited for smaller laboratories and work areas
The Fisherbrand GT05LFEEFSA Undercounter Lab Freezer provides dependable low-temperature storage in a compact design suited for laboratories with limited floor space. Offering a 4.5 cu. ft. (127 L) capacity and a temperature range of -20°C to -10°C, the freezer provides controlled storage for temperature-sensitive laboratory materials. The GT05LFEEFSA features a manual defrost cooling system and a lockable door for added security, while its undercounter configuration allows it to integrate efficiently into laboratory workspaces.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Fisherbrand GT05LFEEFSA Undercounter Lab Freezer includes:
- 2 x Shelves
- Power Cord
- Full 180 Day Parts and Labor Warranty
Fisherbrand GT05LFEEFSA Undercounter Lab Freezer Specifications
| Capacity | 4.5 cu. ft. (127 L) |
| Temperature Range | -20 to -10°C (-4 to 14°F) |
| Defrost Type | Manual |
| Door Lock | Yes |
| Dimensions (W x D x H) | 23 x 22.5 x 34 in. (58.4 x 57.2 x 86.4 cm) |
| Electrical Requirements | 115V, 60Hz, 1.1A |
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for pixel issues and replace if needed | Completed | |
| Test button functionality and replace any defective buttons | Completed | |
| Examine door seal and lock functionality; replace if compromised | Completed | |
| Conduct PCB test point diagnostics and replace as needed | Completed | |
| Verify compressor and internal fan functionality | Completed | |
| Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
| Inspect all wiring for frays, tears, or breakages | Completed | |
| Test start capacitor capacitance range; replace if needed | Completed | |
| Replace all internal filters and clean condenser fins | Completed | |
| Test alarm battery for full functionality | Completed | |
| Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
| Conduct cosmetic inspection and address any issues | Completed |