Fisher Scientific 14955241 Mini Heat Block
Temperature Range
Ambient +5°C to 100°C
Temperature Accuracy
± 0.3°C
Temperature Uniformity
± 0.2°C
- Compact dry bath design ideal for space-limited laboratory environments
- Dual-heating system with independent lid and block elements ensures uniform heat distribution
- Minimizes sample condensation and evaporation for improved sample integrity
- Independently adjustable lid and block temperatures in precise 0.1°C increments
- Suitable for sensitive applications such as LAMP assays and immunoassays
- Reliable performance in clinical, research, and diagnostic labs
The Fisher Scientific 14955241 Mini Heat Block is a high-precision, compact dry bath block heater designed with advanced temperature control features. This mini dry bath incorporates a unique dual-heating system, consisting of a main heating element and a second one within its lid, to ensure uniform temperature distribution throughout the sample. This design effectively reduces condensation and evaporation, maintaining sample integrity. The device offers precise temperature control, with both block and lid temperatures independently adjustable in 0.1°C increments. This level of precision, combined with a temperature range from Ambient +5°C up to 100°C, temperature accuracy of ± 0.3°C, and uniformity of ± 0.2°C, makes it ideal for temperature-sensitive applications such as LAMP assays and immunoassay procedures. The Fisher Scientific 14955241 Mini Heat Block is a reliable and efficient choice for laboratory professionals seeking consistent and accurate heating.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
The Fisher Scientific 14955241 Mini Heat Block includes:
- 15 x 1.5ml Mini Block
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Fisher Scientific 14955241 Mini Heat Block Specifications
| Temperature Range | Ambient +5°C to 100°C |
| Temperature Accuracy | ± 0.3°C |
| Temperature Uniformity | ± 0.2°C |
| Weight | 1.8 lb. (0.8 kg.) |
| Electrical Requirements | 100-240V, 2A (24V Power Supply) |
- Verify the number and correct sizes of heat blocks to ensure compliance with system specifications
- Test functionality of all displays, knobs, and switches, ensuring reliable control and accurate readings
- Inspect all internal wiring for any signs of damage, including frays or compromised insulation
- Clean internal insulation and chamber to maintain optimal thermal efficiency and system hygiene
- Check machine feet and confirm the unit sits properly balanced for stable operation
- Use an independent temperature probe to confirm heating accuracy within the specified temperature range
- Validate temperature uniformity across heat blocks for consistent thermal distribution
- Inspect line cord for any signs of wear, cuts, or damage affecting safety or performance
- Ensure cosmetic condition meets quality standards, free of damage that could impact operation or aesthetics
Technician Inspection and Recertification Checklist: Dry Bath Heater
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Verify the number and sizes of heat blocks | Completed | |
| Test functionality of displays, knobs, and switches | Completed | |
| Inspect internal wiring for damage | Completed | |
| Clean internal insulation and chamber | Completed | |
| Check machine feet for proper balance | Completed | |
| Confirm heat within range using an independent probe | Completed | |
| Validate temperature uniformity across heat blocks | Completed | |
| Inspect line cord for damage | Completed | |
| Ensure cosmetic condition is acceptable | Completed |