Binder BD56 Avantgarde Gravity Convection Incubator, Cat #: 9010-0324
Volume
56.6 L (2 cu. ft.)
Temperature Range
+5°C above ambient to 100°C
Temperature Variation
±0.3 K at 37°C
- Gentle gravity convection heating for uniform temperature distribution
- Easy USB data logging with standard USB interface
- Intuitive LCD controller for simple operation
- Programmable ramp function for controlled heating profiles
- Delayed ON and delayed OFF timer functions
- Class 3.1 independent temperature safety device for enhanced sample protection
- Visual temperature alarm for monitoring operating conditions
- Optimized heating system with low heat dissipation for outstanding energy efficiency
- Low installation requirements reduce operating costs and simplify placement
- Rapid heating to operating temperature with fast recovery following door openings
- Supports up to four shelves with two standard shelves included
- High load capacity with up to 154 lb permitted chamber load and 66 lb per shelf
- Compact benchtop design ideal for research, pharmaceutical, clinical, and quality control laboratories

The Binder BD56 Avantgarde Gravity Convection Incubator, Cat #: 9010-0324 is designed to provide dependable, energy-efficient incubation for microbiology, pharmaceutical, clinical, and research laboratories. Utilizing gravity convection technology, the BD56 delivers gentle, uniform heating that is ideal for routine incubation of sensitive samples while maintaining excellent temperature stability and consistency. With an operating temperature range from 5°C above ambient to 100°C, the incubator offers the flexibility required for a wide variety of laboratory applications, while its optimized heating system minimizes heat dissipation and reduces installation requirements for improved energy efficiency.
Engineered for ease of use, the BD56 features an intuitive LCD controller with programmable ramp and timer functions, allowing users to easily configure temperature profiles and automated start and stop times. A standard USB interface enables convenient data logging and simplifies documentation of incubation conditions. The incubator also incorporates a Class 3.1 independent temperature safety device and a visual temperature alarm to help protect valuable samples throughout operation. Exceptional thermal performance is achieved through temperature uniformity of ±0.3 K at 37°C and ±1.1 K at 100°C, with fluctuations as low as ±0.1 K at 37°C, ensuring accurate, repeatable results for demanding laboratory workflows.
The Binder BD56 Avantgarde Gravity Convection Incubator, Cat #: 9010-0324 includes:
- 2 x Shelves
- NIST Calibration and Preventative Maintenance Report
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Binder BD56 Avantgarde Gravity Convection Incubator, Cat #: 9010-0324 Specifications
| Volume | 56.6 L (2 cu. ft.) |
| Temperature Range | +5°C above ambient to 100°C |
| Temperature Variation | ±0.3 K at 37°C |
| Temperature Fluctuation | ±0.1 K at 37°C |
| Heating-Up Time | 45 min to 37°C |
| Recovery Time | 16 min (after 30-second door opening at 37°C) |
| Maximum Load per Rack | 66 lb (30 kg) |
| Interior Dimensions (W x D x H) | 14.1 x 14.9 x 16.5 in. (36.0 × 38.0 × 42.0 cm) |
| Exterior Dimensions (W x D x H) | 22.1 x 22.3 x 24.6 in. (56.0 x 56.5 x 62.6 cm) |
| Weight | 86 lb. (39 kg) |
| Electrical Requirements | 120V, 50/60 Hz |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and CO₂ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |
Downloads


Click Here to View the Technical Specs for the Binder BD56 Avantgarde Gravity Convection Incubator, Cat #: 9010-0324