Benchmark MyTemp 65HC Digital Cooling Incubator
The Benchmark MyTemp 65HC Digital Cooling Incubator is designed to provide accurate temperature control from Ambient -25°C to 60°C with a uniformity of ± 1.0°C at 37°C. The broad temperature range of the instrument expands the use of this incubator to both heating and cooling applications. A large, digital display allows for visualization of the chamber temperature and the remaining incubation time, while an internal light is included for improved visualization of the chamber. In addition, a built-in internal outlet can be used to provide power to small instruments, such as shakers, rockers, etc. With forced air technology and stainless-steel inner construction, uniform heat transfer is ensured throughout the internal chamber. When set for sub-ambient temperatures, the powerful compressor is activated, quickly cooling the chamber to the desired temperature. With an exterior height of 39.35 inches, this incubator is designed to be floor-standing, yet is also compact enough for bench-use.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Benchmark MyTemp 65HC Digital Cooling Incubator includes:
- 2 Adjustable Shelves
- NIST Calibration and Preventative Maintenance Report
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Benchmark MyTemp 65HC Digital Cooling Incubator Specifications
| Temperature Range | Ambient -25°C to 60°C |
| Temperature Uniformity | ± 1.0°C at 37°C |
| Temperature Accuracy | 0.5°C at 37°C |
| Temperature Increments | ± 0.1°C |
| Volume | 2.3 cu. ft. (65L) |
| Refrigerant | R134A |
| Internal Dimensions (W x D x H) | 15.75 x 12.75 x 19.75 in. (40 x 32.4 x 50.2 cm) |
| External Dimensions (W x D x H) | 21.25 x 21 x 39.35 in. (54 x 53.34 x 99.85 cm) |
| Weight | 117 lb. (53 kg.) |
| Electrical Requirements | 120V, 50/60 Hz, 385W, 3.2A |
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Inspect display for dim, flickering, or dead pixels; replace for readability | Completed | |
| Examine external regulators for damage or malfunction; replace compromised parts | Completed | |
| Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy | Completed | |
| Inspect internal wiring for frays, tears, or breakages; replace compromised wires | Completed | |
| Verify software settings align with components and calibration specs; adjust if needed | Completed | |
| Clean fan and remove dust/obstructions; replace if performance is substandard | Completed | |
| Examine wheels and leveling mounts for structural integrity and balance | Completed | |
| Inspect door gaskets and latches for wear or damage; replace to maintain seal | Completed | |
| Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy | Completed | |
| Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision | Completed | |
| Monitor temperature and CO₂ stability over 24-hour period to verify consistency | Completed | |
| Conduct cosmetic inspection to confirm quality of external surfaces and components | Completed |



Click Here to View the Technical Specs for the Benchmark MyTemp 65HC Digital Cooling Incubator