Beckman Allegra 6 Centrifuge
Maximum Speed
6,100 RPM (3,750 RPM with included rotor)
Maximum Capacity
4 x 750ml
Timer
30 Minutes
- Smart Balance Technology ensures accurate performance with automatic imbalance detection and shutdown for safety.
- Features an electromechanical door-locking mechanism with a secondary manual lock, enhancing operator safety.
- Monitors rotor load during operation, automatically shutting down in case of severe imbalance.
- Equipped with a rugged drive and fine-tuned suspension for consistent performance and reliability in the lab.
- Two independent monitoring systems prevent door opening when the rotor is spinning, ensuring enhanced lab safety.
- Includes anchoring provisions to secure the centrifuge, minimizing risk of injury or damage in case of rotor issues.

The Beckman Allegra 6 Centrifuge features smart balance technology and has a high capacity with the GH-3.8 swing bucket rotor. This benchtop centrifuge and its rotor can accommodate a wide range of vessels, from 1.5 ml tubes to 750 ml bottles and generally including tubes, bottles, microplates, specialty bottles, and blood bags. The Allegra 6 reaches maximum speeds up to 6,100 RPM with maximum RCF of 5,710 x g. Features automatic rotor imbalance detection, auto door interlock, rugged drive, and fine-tuned suspension for performance, reliability, and safety in the lab.
The Allegra 6 is ideal for cell culture, plasma, and other general purpose separations.
All of our equipment is fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Beckman Coulter Allegra 6 Centrifuge includes:
- GH-3.8 Swing Bucket Rotor and Buckets
- 1 Set of 15ml Adapters
- 1 Set of 50ml Adapters
- Power Cord
- Operator's Manual
- NIST Centrifuge Calibration and Preventative Maintenance Report
- Full 180 Day Parts and Labor Warranty
Beckman Allegra 6 Centrifuge Specifications
|
Maximum speed (RPM) |
6,100 RPM (3,750 RPM with included rotor) |
|
Maximum capacity |
4 x 750ml |
|
Run time |
0 – 30 minute, hold timer |
|
Driver |
Direct drive with replacement brushes |
|
Accel/Decel rates |
1 / 3 |
|
Maximum noise output |
< 60 dB |
|
Dimensions (W x D x H) |
21.5 x 26 x 15.5 in |
|
Weight |
114 lbs |
| Refrigeration | No Refrigeration |
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply output voltages match specified requirements | Completed | |
| Measure and verify voltages at PCB test points for accuracy | Completed | |
| Confirm rotor compatibility and detection for proper identification | Completed | |
| Inspect exterior for cosmetic or structural damage | Completed | |
| Check door mechanism alignment for secure opening and closing | Completed | |
| Inspect and replace gas strut with OEM part as needed | Completed | |
| Examine rotor chamber for dents or damage and replace if necessary | Completed | |
| Replace and confirm functionality of all lid lock switches | Completed | |
| Lubricate door lock mechanism and lid lock rail for smooth operation | Completed | |
| Test compressor (if applicable) for proper operation and cooling | Completed | |
| Run 8-12 hour cycle test to ensure consistent performance | Completed | |
| Check rotor motor for stability, noise, and integrity | Completed | |
| Inspect and clear condensation drain to prevent buildup | Completed | |
| Verify RPM accuracy with NIST traceable tachometer and calibrate | Completed | |
| Check centrifuge error log for zero recorded issues | Completed |
Preventative Maintenance Performed on The Beckman Coulter Allegra 6 Centrifuge
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
Beckman Allegra 6 Centrifuge Manual

Click Here to View all the Technical Specs for the Allegra 6R




Click Here to View the Technical Specs for the Allegra 6R