Wavelength Range
190 - 600 nm
Noise
< ± 0.5 x 10-5 AU, at 254 nm
Drift
3 x 10-4 AU/hr, at 254 nm
- Double-beam optical system delivers stable, high-precision detection for reliable analytical results
- High-intensity deuterium lamp supports strong signal output and sensitive measurement performance
- Fast 55 Hz data acquisition enables accurate tracking of narrow peaks in high-speed separations
- Built-in holmium oxide filter allows quick and dependable wavelength accuracy verification
- GLP-ready functionality supports compliance, traceability, and audit-ready laboratory workflows
- Front-access design simplifies routine maintenance with quick lamp and flow cell replacement

The Agilent 1260 Infinity I HPLC G1314C VWD VL+ is engineered to deliver reliable, high-performance detection for demanding HPLC workflows. Designed with a focus on optical precision and operational efficiency, this variable wavelength detector provides stable, accurate measurements across a broad range of applications. At its core, a high-intensity deuterium lamp ensures strong signal output and low detection limits, supporting sensitive analyses while maintaining consistent performance. The system’s fast acquisition capability, operating at 55 Hz, enables accurate tracking of narrow peaks and high-throughput separations, making it well suited for modern fast HPLC methods.
To support data integrity and compliance requirements, the detector includes a built-in holmium oxide filter for quick and reliable wavelength accuracy verification. Routine maintenance is streamlined through convenient front access to key components such as the lamp and flow cell, allowing for quick replacement and minimal downtime. With its combination of robust optical performance, compliance-ready features, and user-friendly serviceability, the G1314C VWD VL+ is a dependable solution for laboratories seeking consistent, high-quality detection in everyday analytical workflows.
The Agilent 1260 Infinity I HPLC G1314C VL+ VWD Detector includes:
- Firmware Version 07.02 (Can be changed upon request)
- Flow Cell
- Deuterium Lamp
- Lab Advisor Calibration Report
- Lan Cable / Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Agilent 1260 Infinity I HPLC G1314C VL+ VWD Detector Specifications
| Detection Type | Double-beam photometer |
| Light Source | Deuterium lamp |
| Wavelength Range | 190 - 600 nm |
| Noise | < ± 0.5 x 10-5 AU, at 254 nm |
| Drift | 3 x 10-4 AU/hr, at 254 nm |
| Maximum Data Rate | 55 Hz |
| Linearity | > 2 AU (5%) upper limit |
| Wavelength Accuracy | ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter |
| Wavelength Precision | < ± 0.1 nm |
| Slit Width | 6.5 nm typical over whole wavelength range |
| Time Programmable | Wavelength, polarity, peak width, lamp on/off |
| Flow Cell | Standard: 14 ul volume, 10 mm cell path length and 40 bar (580 psi) pressure maximum |
| Spectral Tools | Stop-flow wavelength scan |
| Dimensions (H x W x D) | 5.5 x 13.5 x 17 in. (14 x 34.5 x 43.6 cm) |
| Weight | 24.3 lb. (11 kg) |
| Electrical Requirements | 100-240V, 50/60Hz, 80VA |
Downloads