Agilent 1200 HPLC G1329A Autosampler
Injection Range
0.1 – 100 µl in 0.1 µl increments
Precision
Typically < 0.5 % RSD of peak areas from 5 – 100 µl
Typically < 1 % RSD 1 – 5 µl variable volume
Sample Capacity
100 × 2-ml vials in 1 tray
40 × 2-ml vials in ½ tray
15 × 6-ml vials in ½ tray (Agilent vials only)
- Equipped with GLP features, including early maintenance feedback (EMF) and electronic records of maintenance and errors
- Multiple communication options with CAN, GPIB, RS232C, and APG-remote interfaces for seamless integration
- Advanced safety features such as leak detection, safe leak handling, and low voltage in maintenance areas
- Error detection and display system for quick issue identification and resolution
- Supports 1–99 replicate injections from a single vial for enhanced workflow efficiency
- Low carryover rate, typically < 0.1%, further reduced to < 0.05% with external needle cleaning
- Handles a wide sample viscosity range from 0.2 to 50 cp for diverse sample types
- Fast injection cycle time, typically 50 seconds, for increased throughput

The Agilent 1200 HPLC G1329A Autosampler is a variable volume autosampler with a sample capacity of 100 samples. The G1329A's variable volume injections range from 1 to 100 µl and are programmable and temperature controlled. The temperature ranges from 4° to 40°C with a temperature accuracy of -1° to +5°C.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Agilent 1200 HPLC G1329A Autosampler includes:
- Latest Available Firmware Version (Please inquire)
- Operator's Guide
- 100 Vial Tray
- Front Cover
- Lab Advisor Documentation
- Power Cord
- Full 180 Day Parts and Labor Warranty
Agilent 1200 HPLC G1329A Autosampler Specifications
|
GLP Features |
Early maintenance feedback (EMF), electronic records of maintenance and errors |
|
Communications |
Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output |
|
Safety Features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
|
Injection Range |
0.1 – 100 µl in 0.1 µl increments Up to 1500 µl with multiple draw (hardware modification required) |
|
Replicate Injections |
1 – 99 from one vial |
|
Precision |
Typically < 0.5 % RSD of peak areas from 5 – 100 µl Typically < 1 % RSD 1 – 5 µl variable volume |
|
Minimum Sample Volume |
1 µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvial |
|
Carryover |
Typically < 0.1 %, < 0.05 % with external needle cleaning |
|
Sample Viscosity Range |
0.2 – 50 cp |
|
Replicate Injections Per Vial |
1 – 99 |
|
Sample Capacity |
100 × 2-ml vials in 1 tray 40 × 2-ml vials in ½ tray 15 × 6-ml vials in ½ tray (Agilent vials only) |
|
Injection Cycle Time |
Typically 50 s depending on draw speed and injection volume |



Click Here to View the Technical Specs for the G1329A