Labconco 4612021 ScrubAir Pipette Washer & Dryer w/ Heater
Maximum Fill Water Temperature
60°C (140°F)
Pump Rating
96 0.46 gal (1.75l) to 0.7 gal (2.5l)
Heater Power
750W
- Automated system washes, rinses, and dries volumetric pipettes efficiently
- Integrated heater supports improved cleaning performance and faster drying
- Graphic display shows cycle status and remaining time for easy monitoring
- Transparent acrylic cover allows visual observation during operation
- Programmable wash and rinse cycles support customized cleaning routines
- Multiple rinse cycles with adjustable durations enhance cleaning flexibility
- Delayed start function enables unattended operation
- Compatible with a wide range of pipette sizes for versatile use
- Uses water and compressed air for thorough and consistent results
- User-friendly controls simplify setup and operation
A transparent acrylic cover enables visual observation of the process, while flexible programming options allow users to select pre-set cycles or create customized wash routines to meet specific requirements. Adjustable wash and rinse parameters, along with delayed start functionality, provide added convenience for unattended operation. Designed for ease of use and adaptability, the Labconco 4612021 ScrubAir Pipette Washer & Dryer offers a reliable and user-friendly solution for maintaining clean, ready-to-use pipettes in laboratory environments.
Please Note: The Labconco 4612021 ScrubAir Pipette Washer should be attached to a regulated source of air, or other suitable inert gas. Pressure must be at least 20 psi (138kPa) and a minimum flow rate of 5 cfm (142 l/m). The pressure must not exceed 100 psi (689kPa).
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Labconco 4612021 ScrubAir Pipette Washer & Dryer w/ Heater includes:
- Serological Pipette Holder
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Labconco 4612021 ScrubAir Pipette Washer & Dryer w/ Heater Specifications
| Minimum Fill Water Temperature | 49°C (120°F) |
| Maximum Fill Water Temperature | 60°C (140°F) |
| Heater Power | 750W |
| Water Consumption per Fill | 0.46 gal (1.75l) to 0.7 gal (2.5l) |
| Minimum Feed Water Pressure | 8 psi (55 kPa) |
| Maximum Feed Water Pressure | 120 psi (827 kPa) |
| Minimum Feed Air Flow / Pressure | 5 cfm (142 L/min), 20 psi (138 kPa) |
| Maximum Feed Air Pressure | 100 psi (689 kPa) |
| Dimensions (W x D x H) | 14.8 x 18.6 x 16.8 in. (37.6 x 47.3 x 42.7 cm) |
| Weight | 84 lb. (38 kg) |
| Electrical Requirements | 115V, 50/60Hz, 2A |
- Inspect the display for pixel clarity and brightness issues; replace any dim or defective pixels for optimal performance.
- Test each button for responsiveness, resistance, and continuity, replacing any that do not meet operational standards.
- Verify touchscreen functionality (if applicable), ensuring consistent response across all touch points.
- Check the door seal and lock for airtight closure and secure operation, replacing any compromised parts.
- Inspect rubber hoses for wear, tears, or cracks, replacing any that compromise system integrity.
- Inspect for internal or external leakage, addressing and sealing any detected points of vulnerability.
- Check all wiring for integrity, inspecting for frays, tears, or breakages, and replace compromised sections.
- Run the system in continuous cycle mode, monitoring for operational errors to ensure reliable performance.
- Confirm all sensors are activated and responsive using internal diagnostics, recalibrating or replacing any faulty ones.
- Conduct a cosmetic inspection of external surfaces, verifying cleanliness and addressing visible wear to meet standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect display for pixel issues and replace if needed | Completed | |
| Test button functionality and replace any defective buttons | Completed | |
| Verify touchscreen functionality (if applicable) | Completed | |
| Check door seal and lock functionality; replace if compromised | Completed | |
| Inspect rubber hoses for wear, tears, or cracks | Completed | |
| Check for internal or external leakage and address as needed | Completed | |
| Inspect wiring for frays, tears, or breakages | Completed | |
| Run continuous cycle mode and monitor for errors | Completed | |
| Confirm all sensors are activated and functional | Completed | |
| Conduct cosmetic inspection and address visible wear | Completed |

Click Here to View the Technical Specs for the Labconco 4612021 ScrubAir Pipette Washer & Dryer w/ Heater