Corning PC-600 Hot Plate
The Corning PC-600 Hot Plate delivers reliable and precise heating, making it an essential tool for any laboratory. With a temperature range of 25°C to 550°C, it provides consistent performance for a wide variety of applications. The 10 x 10-inch ceramic platform features Corning’s exclusive Pyroceram® surface, offering superior resistance to scratches, corrosion, and chemical attack. Its smooth, easy-to-clean design ensures uniform heating and excellent durability, while the bright white surface enhances visibility for procedures requiring close monitoring.
Designed with user safety and convenience in mind, the PC-600 includes enclosed control electronics that protect sensitive components from harsh environments, reducing maintenance time and costs. Tested to rigorous electrical safety and regulatory standards, it meets UL, CUL, IEC1010, CE, GS, and FCC requirements for dependable, worry-free operation. Compact yet robust, this hot plate measures 4.8 x 10.5 x 14.4 inches and weighs 10 lbs., making it ideal for both bench-top efficiency and portability. Powered by a 120V system with 1043W, the PC-600 combines performance, safety, and reliability in one trusted device.
All of our equipment is in excellent condition and fully tested before shipping.
The Corning PC-600 Hot Plate includes:
- Power Cord
- Operator's Manual
- Full 180 Day Parts and Labor Warranty
Corning PC-600 Hot Plate Specifications
| Temperature Range | 25°C to 550°C |
| Platform Size | 10 x 10 in (25.4 x 25.4 cm) |
| Platform Material | Ceramic |
| Dimensions (H x W x D) | 4.8 x 10.5 x 14.4 in. (12.2 x 26.7 x 36.6 cm) |
| Weight | 10.0 lb. (4.5 kg) |
| Electrical Requirements | 120V, 60Hz, 8.7A, 1043W |
- Inspect Display and Replace Any Dim/Dead Pixels (If Applicable).
- Inspect Ceramic Top for Any Cracks and Replace If Necessary.
- Inspect Functionality of All Buttons/Knobs.
- Use Magnetic Stir Bar Alongside NIST Tachometer to Confirm RPM of Internal Motor (If Applicable).
- Inspect Internal Wiring Inside Case.
- Clean Internal Parts and Vent Any Internal Debris.
- Check Feet to Confirm Plate Sits Properly Balanced.
- Replace All Parts That Were Removed on Internal Inspections.
- Confirm Temperature with NIST Traceable Temperature Probe (If Applicable).
- Ensure Cosmetics Are Acceptable.
Technician Inspection and Recertification Checklist: Hot Plate Stirrers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment| Task | Completed? | Remarks? |
| Inspect Display and Replace Any Dim/Dead Pixels (If Applicable) | Completed | |
| Inspect Ceramic Top for Any Cracks and Replace If Necessary | Completed | |
| Use Magnetic Stir Bar Alongside NIST Tachometer to Confirm RPM of Internal Motor (If Applicable) | Completed | |
| Inspect Internal Wiring Inside Case | Completed | |
| Clean Internal Parts and Vent Any Internal Debris | Completed | |
| Check Feet to Confirm Plate Sits Properly Balanced | Completed | |
| Replace All Parts That Were Removed on Internal Inspections | Completed | |
| Confirm Temperature with NIST Traceable Temperature Probe (If Applicable) | Completed |


Click Here to View the Technical Specs for the Corning PC-600 Hot Plate