Thermo Scientific Precision DUB 15 Shaking Water Bath
The Thermo Scientific Precision DUB 15 Shaking Water Bath supports a range of sensitive life science and QA/QC applications from warming fragile reagents to tissue culturing and genetics sequencing. The DUB 15, 15L water bath is easy to clean and maintain with coil-free design. Conveniently save commonly used settings with four temperature and shaking speed presets while optimizing scheduling with the featured auto-on and auto off timers.
All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
> Bath Lid
> Power Cord
> Operator's Guide
> Full 180 Day Parts and Labor Warranty
Thermo Scientific Precision DUB 15 Shaking Water Bath Specifications
Temperature Range |
Ambient +5C to 100C |
Heating Capacity |
1200W |
Volume |
4 gal, 15L |
Control |
Digital |
Electrical Requirements |
Global Voltage 100-115V/200-230V, 50/60Hz |
Dimensions (L x W x H) Exterior |
25 x 15.6 x 9.9 in. (636 x 396 x 250 mm) |
Dimensions (D x W x H) Interior |
12 x 15.3 x 7.6 in. (305 x 387 x 192 mm) |
Amperage |
9.0-10.5 / 4.3-5.0 (120V/230V) |
Temperature Stability |
0.1 C |
Temperature Uniformity |
0.05 C |
Temperature Presets |
4 |
- Confirm power supply and PCB are outputting stable, correct voltages across all essential circuits to ensure reliable operation
- Inspect the display for any dim, flickering, or dead pixels, replacing any compromised sections to maintain full clarity and accurate readouts
- Verify that all buttons, knobs, and switches are functioning properly, providing precise and responsive control across all settings
- Conduct a thorough inspection of the water reservoir for any holes, rust, or other structural damage that may impact its integrity
- Fill the reservoir with water, checking closely for any external or internal leaks that could affect performance
- Examine the lid for any signs of wear or damage, replacing hinges and handles if necessary to ensure a secure and proper seal
- Inspect for any internal leaks originating from the water reservoir, confirming the unit’s internal plumbing integrity
- Conduct a comprehensive inspection of internal wiring for any breakages, tears, or wear; replace any underperforming or damaged components to maintain safe operation
- Examine internal insulation for signs of burns, breaks, or deterioration, replacing as needed to ensure thermal efficiency and safety
- Confirm that all feet are intact and that the bath sits level and balanced on a flat surface to prevent movement or instability
- Calibrate both temperature accuracy and uniformity using a NIST-traceable temperature probe, verifying compliance with specification standards
- Perform a detailed cosmetic inspection to confirm that all external surfaces and finishes meet aesthetic and quality standards
Technician Inspection and Recertification Checklist: Water Bath
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment| Task | Completed? | Remarks? |
| Confirm power supply and PCB output stable, correct voltages across circuits | Completed | |
| Inspect display for dim, flickering, or dead pixels; replace for clarity | Completed | |
| Verify functionality of all buttons, knobs, and switches for responsive control | Completed | |
| Inspect water reservoir for holes, rust, or structural damage | Completed | |
| Fill reservoir with water; check for external or internal leaks | Completed | |
| Examine lid for wear or damage; replace hinges or handles as needed | Completed | |
| Inspect for internal leaks from water reservoir to confirm plumbing integrity | Completed | |
| Check internal wiring for damage; replace underperforming components for safety | Completed | |
| Examine internal insulation for burns, breaks, or deterioration; replace as needed | Completed | |
| Confirm feet are intact and bath is balanced on flat surface | Completed | |
| Calibrate temperature accuracy and uniformity with NIST-traceable probe | Completed | |
| Perform cosmetic inspection to confirm quality of external surfaces and finishes | Completed |