|
|
|
|
Corning PC-600 Hot Plate
The Corning PC-600 Hot Plate delivers reliable and precise heating, making it an essential tool for any laboratory. With a temperature range of 25°C to 550°C, it provides consistent performance for a wide variety of applications. The 10 x 10-inch ceramic platform features Corning’s exclusive Pyroceram® surface, offering superior resistance to scratches, corrosion, and chemical attack. Its smooth, easy-to-clean design ensures uniform heating and excellent durability, while the bright white surface enhances visibility for procedures requiring close monitoring.
Designed with user safety and convenience in mind, the PC-600 includes enclosed control electronics that protect sensitive components from harsh environments, reducing maintenance time and costs. Tested to rigorous electrical safety and regulatory standards, it meets UL, CUL, IEC1010, CE, GS, and FCC requirements for dependable, worry-free operation. Compact yet robust, this hot plate measures 4.8 x 10.5 x 14.4 inches and weighs 10 lbs., making it ideal for both bench-top efficiency and portability. Powered by a 120V system with 1043W, the PC-600 combines performance, safety, and reliability in one trusted device. All of our equipment is in excellent condition and fully tested before shipping.
The Corning PC-600 Hot Plate includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Corning PC-600 Hot Plate Specifications
Temperature Range
|
25°C to 550°C
|
Platform Size
|
10 x 10 in (25.4 x 25.4 cm)
|
Platform Material
|
Ceramic
|
Dimensions (H x W x D)
|
4.8 x 10.5 x 14.4 in. (12.2 x 26.7 x 36.6 cm)
|
Weight
|
10.0 lb. (4.5 kg)
|
Electrical Requirements
|
120V, 60Hz, 8.7A, 1043W
|
- Inspect the display for clarity and brightness, replacing any dim or dead pixels to ensure optimal performance.
- Examine external regulators for structural integrity and damage, replacing if necessary (if applicable).
- Assess the functionality of temperature, CO₂, and H₂O sensors, verifying accuracy and consistency (if applicable).
- Check all internal wiring for frays, tears, or breakages, addressing any compromised sections.
- Confirm software settings are properly configured to match the specifications of internal components.
- Inspect and clean the fan for efficient airflow, replacing if wear or malfunction is detected.
- Verify the condition of wheels and leveling mounts, ensuring stability and ease of movement.
- Inspect door gaskets and latches for sealing integrity, replacing if compromised to maintain proper enclosure.
- Calibrate temperature accuracy to 37°C using an NIST-traceable temperature probe for precision.
- Calibrate CO₂ levels to 5% using a Fyrite combustion test kit to verify environmental control (if applicable).
- Confirm temperature and CO₂ stability over a 24-hour period to ensure consistent conditions.
- Conduct a cosmetic inspection, verifying the unit meets appearance and cleanliness standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for clarity and replace if needed
|
Completed
|
|
Examine external regulators for integrity and damage
|
Completed
|
|
Assess temperature, CO₂, and H₂O sensors for accuracy
|
Completed
|
|
Check internal wiring for frays, tears, or breakages
|
Completed
|
|
Confirm software settings match internal component specs
|
Completed
|
|
Inspect and clean fan, replacing if wear is detected
|
Completed
|
|
Verify condition of wheels and leveling mounts
|
Completed
|
|
Inspect door gaskets and latches for sealing integrity
|
Completed
|
|
Calibrate temperature to 37°C with NIST-traceable probe
|
Completed
|
|
Calibrate CO₂ to 5% with Fyrite combustion test kit
|
Completed
|
|
Confirm 24-hour temperature and CO₂ stability
|
Completed
|
|
Conduct cosmetic inspection to ensure appearance standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|