BioTek Synergy MX Multi-Mode Microplate Reader
Detection Modes Absorbance, Fluorescence Intensity, Luminescence
Microplate Formats 1- to 384- well
Read Types Endpoint, Kinetic, Spectral Scanning, Well-Area Scanning
- Fully monochromator-based with quadruple grating architecture for ultra fine-tuned performance
- Compatible with Take3 Multi-Volume Plate with 2 µL microspots for low volume assays
- High performance fluorescence (top read only), absorbance and dedicated luminescence detection system accommodates all life-science research microplate applications
- Automated Z-height adjustment accommodates variable microplate heights and liquid levels with 100 μm resolution
- Variable bandpass selection system adds flexibility
- Cuvette port to measure standard 1-cm path cuvettes
- Controlled by the powerful BioTek Gen5 software, facilitating easy data collection, analysis, exporting, and reporting
The BioTek Synergy MX Multi-Mode Microplate Reader is an advanced, versatile laboratory instrument designed for various life-science research applications. This state-of-the-art device features a fully monochromator-based system with quadruple grating architecture, ensuring ultra-fine-tuned performance across a range of functions. It is compatible with the Take3 Multi-Volume Plate, which includes 2 µL microspots for conducting low volume assays efficiently. Key features of the Synergy MX include high-performance fluorescence detection (top read only), absorbance, and a dedicated luminescence detection system. Its automated Z-height adjustment can accommodate varying microplate heights and liquid levels with a precise 100 μm resolution. The variable bandpass selection system adds further flexibility to its operations. Additionally, the device includes a cuvette port for standard 1-cm path cuvettes.
The absorbance specifications of the Synergy MX are impressive. It utilizes a Xenon flash lamp as its light source and a double monochromator for wavelength selection, covering a range from 230 to 999 nm in 1 nm increments. The device ensures photometric accuracy of < 1% at 2.0 OD and repeatability of < 0.5% at 2.0 OD, with a dynamic range from 0 to 4.0 OD. For fluorescence intensity measurements, the device also uses a Xenon flash lamp and a dedicated red-shifted PMT detector, with a wavelength range of 250 to 900 nm and a remarkable sensitivity level. The luminescence capabilities of the Synergy MX are equally robust, featuring a low noise PMT detection system and a dynamic range of < 6 decades, with exceptional sensitivity for detecting as low as 10 amol of ATP.
This microplate reader can accommodate formats from 1 to 384-well plates and PCR plates, and offers various read methods such as endpoint, kinetic, spectral scanning, and well-area scanning. It also includes plate shaking features and precise temperature control from Ambient +4°C to 65°C. Controlled by BioTek's powerful Gen5 software, the Synergy MX facilitates easy data collection, analysis, exporting, and reporting, making it an invaluable tool for modern laboratories engaged in life-science research. Please Note: This is the SMATCL model of the Synergy MX, which DOES NOT feature bottom fluorescence reading or a dual reagent dispenser.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The BioTek Synergy MX Multi-Mode Microplate Reader includes:
- > Windows 11 Professional 64-Bit Data Station with Pre-Loaded Gen5 PC Software
- > ABS, FLD, LUM, and Self Test Calibration Report
- > Interface Cable
- > Operator's Manual
- > Full 180 Day Part and Labor Warranty
BioTek Synergy MX Multi-Mode Microplate Reader Specifications
Absorbance Specifications
|
Light Source
|
Xenon flash lamp
|
Wavelength Selection
|
Double monochromator
|
Wavelength Range
|
230 to 999 nm, in 1 nm increments
|
Bandpass
|
2 nm (230-285 nm), 4 nm (>285 nm)
|
Pathlength Correction
|
Yes
|
Monochromator Wavelength Accuracy
|
± 2 nm
|
Monochromator Wavelength Repeatability
|
± 0.2 nm
|
Dynamic Range
|
0 to 4.0 OD
|
Photometric Resolution
|
0.0001 OD
|
Photometric Accuracy
|
< 1% at 2.0 OD typical
|
Photometric Linearity
|
< 1% from 0 to 3.0 OD
|
Photometric Repeatability
|
< 0.5% at 2.0 OD
|
Reading Speed
|
96 wells: 11 seconds
384 wells: 22 seconds
|
Fluorescence Intensity Specifications
|
Light Source
|
Xenon flash lamp
|
Detector
|
Dedicated red shifted PMT
|
Wavelength Selection
|
Double grating monochromators (top read only)
|
Wavelength Range
|
250 to 900 nm
|
Dynamic Range
|
5 decades
|
Sensitivity (Top Read Only)
|
Fluorescein 2 pM typical (0.2 fmol/well, 384-well plate)
|
Luminescence Specifications
|
Dynamic Range
|
< 6 decades
|
Sensitivity
|
10 amol ATP typical (flash)
|
Detection System
|
Low noise PMT
|
Wavelength Range
|
300 to 700 nm
|
General Specifications
|
Microplate Formats
|
1- to 384- well plates and PCR plates
|
Read Methods
|
End point, kinetic, spectral scanning, well-area scanning
|
Plate Shaking
|
Yes
|
Temperature Control
|
Ambient +4°C to 65°C, ± 0.5°C at 37°C
|
Dimensions (W x D x H)
|
17 x 20.9 x 15 in. (43.5 x 53.1 x 38.1 cm)
|
Weight
|
70 lb. (31.5 kg.)
|
Electrical Requirements
|
100-240V, 50/60Hz
|
Computer Specifications
|
Processor
|
Intel Core I5-8350U CPU
|
RAM
|
16.0 GB
|
Hard Drive
|
512GB Internal SSD
|
Operating System
|
Windows 11 Professional, 64-bit
|
Frequently Purchased With
- Verify that the PCB and power supply output proper voltages to all internal components.
- Examine the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal readability.
- Inspect all buttons for responsiveness, confirming full functionality and consistency.
- Assess the plate drawer for proper clearance and smooth operation to ensure accurate plate handling.
- Confirm seamless communication between the plate reader and computer software for data integrity.
- Inspect internal wiring and electrical connections for wear or potential faults, addressing any issues found.
- Examine the lamp, replacing it as needed to maintain proper illumination and performance.
- Execute the required validation protocol to confirm system accuracy and compliance.
- Apply all current updates to the data station and operating system to ensure compatibility and stability.
- Conduct a final cosmetic inspection to verify that all external surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Verify PCB and power supply output proper voltages | Completed | |
Check display for brightness, replace dim or dead pixels | Completed | |
Inspect buttons for full functionality and responsiveness | Completed | |
Assess plate drawer clearance for accurate plate handling | Completed | |
Confirm communication between plate reader and software | Completed | |
Inspect internal wiring and electrical connections for wear | Completed | |
Examine lamp and replace as necessary for illumination | Completed | |
Execute validation protocol for system accuracy | Completed | |
Apply all current updates to data station and OS | Completed | |
Conduct cosmetic inspection to meet appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|