Arctiko ULUF 15 -86C Ultra Low Undercounter Freezer
Temperature Setting -40°C / -86°C
Cooling System Single System / Static
- Single compressor technology that ensures easy maintenance, lower heat dissipation, and low energy consumption
- Newly developed noise reduction technology reduces noise level drastically for a more pleasant work environment
- Filter less construction prevents dust from blocking the airflow which reduces the unit’s ability to dissipate heat effectively, minimizing maintenance procedure
- Direct cooling features evaporator coils to cool each compartment directly for maximum uniformity and minimal recovery time
- A heated door frame prevents ice from building up around the gasket securing a stable and uniform temperature inside the freezer
- The user friendly G214 controller is standard and includes different alarms such as visual, acoustic, open door, and more
- An insulated inner door is featured for providing better recovery time after door openings and optimal temperature stability
- 100% HCFC & CFC free
The Arctiko ULUF 15 -86C Ultra Low Undercounter Freezer is designed for optimal performance in laboratory settings. This compact freezer, ideal for undercounter or benchtop use, offers a space-saving design and a minimal noise level, enhancing the work environment. It features a volume of 0.3 cu. ft. (7L) and can be set at -40°C and -86C. Equipped with advanced single compressor technology, it ensures efficient cooling with reduced heat dissipation and lower energy consumption. The unit's filter less design minimizes maintenance by preventing dust accumulation that can block airflow. Additional features include direct cooling with spaced evaporator coils for uniform temperature and quick recovery, a heated door frame to prevent ice build-up, and an insulated inner door for optimal temperature stability. The user-friendly G214 controller includes multiple alarm functions for enhanced safety. 100% free of HCFC and CFC, the ULUF 15 promises reliability and maintenance ease, making it an excellent choice for any laboratory.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Arctiko ULUF 15 -86C Ultra Low Undercounter Freezer includes:
- > Key for Door Lock
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Arctiko ULUF 15 -86C Ultra Low Undercounter Freezer Specifications
Temperature Setting
|
-40°C / -86°C
|
Volume
|
0.3 cu. ft. (7L)
|
Cooling Technology
|
Single System / Static
|
Defrost Method
|
Manual
|
Refrigerant
|
EP88
|
Controller Model
|
G214, 4.3" display with touch buttons
|
Probe Type
|
PT1000
|
Alarm Styles
|
Text, visual, and acoustic
|
Alarm Functions
|
Power failure, adjustable high / low temperature, open door, probe failure, and low battery (72 hour battery backup for alarms)
|
Data Logging / External Connections
|
Voltage free contact for remote alarm, electronic data logger, RS 485 port, USB port for software update and data download, temperature chart recorder (optional)
|
Vacuum Release Port
|
No
|
Internal Light
|
No
|
Fixed Shelves / Adjustable Shelves
|
0 / 0
|
Interior Cabinet Material
|
Stainless steel
|
Exterior Cabinet Material
|
Electrogalvanized steel
|
Foam Material
|
Polyurethane
|
Insulation Thickness
|
4.7 in. (12 cm)
|
Interior Dimensions (W x D x H)
|
5.91 x 5.63 x 12.2 in. (15 x 14.3 x 31 cm)
|
Exterior Dimensions (W x D x H)
|
15.75 x 24.8 x 26.18 in. (40 x 63 x 66.5 cm)
|
Weight
|
139 lb. (63 kg)
|
Electrical Requirements
|
110V, 50/60Hz, 5.6A
|
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for pixel issues and replace if needed | Completed | |
Test button functionality and replace any defective buttons | Completed | |
Examine door seal and lock functionality; replace if compromised | Completed | |
Conduct PCB test point diagnostics and replace as needed | Completed | |
Verify compressor and internal fan functionality | Completed | |
Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
Inspect all wiring for frays, tears, or breakages | Completed | |
Test start capacitor capacitance range; replace if needed | Completed | |
Replace all internal filters and clean condenser fins | Completed | |
Test alarm battery for full functionality | Completed | |
Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
Conduct cosmetic inspection and address any issues | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|