Flow Rate Range
0.001 to 25 ml/min (up to 50 ml/min during column packing)
Flow Rate Accuracy
± 1.2%
Flow Rate Precision
RSD <0.5%
Pressure Range
0 to 20 MPa (2900 psi)
Pressure Limits
Programmable upper and lower
Viscosity Range
0.35 to 10 cP (5 cP above 12.5 ml.min)
Wavelength Range
Absorbance Range
-6 to 6 AU
Resolution
0.001 mAU
Wavelength Reproducibility
Linearity
Drift
≤ | 0.2 mAU | AU/h (2 mm cell at 280 nm)
Noise
< 0.1 mAU
Operating Pressure
0 to 2 MPa
Lamp Operating Time
> 10,000 hours
Flow Cell
Optical path length: 2 mm
Cell volume: 2 ul
Total volume: 30 ul
Conductivity Range
0.01 to 999.9 mS/cm
Accuracy
± 0.01 mS/cm or ± 2%, whichever is greater (within 0.3 to 300 mS/cm)
0 to 5 MPa
Flow Cell Volume
22 ul
Temperature Monitor Range
0°C to 99°C
Temperature Monitor Accuracy
± 1.5°C within 4°C to 45°C
pH Reading Range
0 to 14
± 0.1 pH unit within pH 2 to 12, temperature within ± 3°C from calibration temperature
0 to 0.5 MPa
76 ul
Number of Fractions
Vessel Types
Fraction Volumes
Spillage-Free Mode
Automatic, Drop Sync, or Accumulator
Flammable Liquids
No
Delay Volume (UV-Dispenser Head)
205 ul, 86 ul with optional tubing kit (i.d. 0.25 mm)
Dimensions (W x D x H)
15.4 x 23 x 12.6 (39 x 58.5 x 32 cm)
Weight
46.3 lb. (21 kg)
190 to 700 nm in steps of 1 nm, up to 3 wavelengths
< 0.08 mAU
> 5000 hours
Total volume: 11 ul
0.01 to 150 ml/min (up to 300 ml/min during column packing)
± 1.5%
0 to 5 MPa (725 psi)
0.35 to 5 cP