Zymark TurboVap 96 Evaporator w/ Manifolds
The Zymark TurboVap 96 Evaporator is a high speed, microprocessor-controlled evaporator that provides simultaneous, automated concentration of multiple samples, unattended operation, and overall convenience. This evaporator can simultaneously process one or two 96-well deepwell plates. The Zymark TurboVap 96 utilizes a gas vortex shearing action to maintain high evaporation rates under mild thermal conditions. When closing the cover, 96 gas nozzles in the manifold align with each well of the microplate or deepwell plate. A helical flow of air is created by the stream of gas directed into each well. The helical flow sets up a vortexing action that provides for sample homogeneity and continuous rinsing of the well wall. The Zymark TurboVap 96 features key applications such as: genomic and proteomic applications that require concentration of purified extracts, evaporation of DMSO solvent in plates as well as HPLC solvent mixtures of Methanol or Acetonitrile with water, pharmaceutical SPE methods, and more. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The Zymark TurboVap 96 Evaporator w/ Manifolds includes:
- > Gas Manifolds
- > Deepwell Plate Mounts
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Zymark TurboVap 96 Evaporator w/ Manifolds Specifications
Capacity
|
2 x 96-Well Deepwell Plates
|
Gas Flow Rate
|
5 to 99 standard cubic feet/hour (SCFH)
|
Temperature Range
|
Ambient to 80C
|
Gas Regulator and Gauge Range
|
30 to 80 psi
|
Timer Range
|
1 to 99 minutes with a one-second resolution
|
Overall Dimensions (H x W x D)
|
12 x 23.5 x 12.5 in (30.5 x 59.7 x 31.8 cm)
|
Electrical Requirements
|
100-120V, 60Hz
|
Zymark TurboVap 96 Evaporator Manual
Click Here to View the Technical Specs for the Zymark TurboVap 96 Evaporator
- Verify internal power supply output for correct voltage to ensure stable and accurate performance
- Confirm fuse holder is secured firmly, preventing movement or disconnection
- Inspect glass window for any cracks or chips, maintaining structural integrity and visibility
- Assess power-up diagnostics, display, and push wheel buttons for full functionality
- Ensure pressure gauge and regulator are operating accurately and within specified ranges
- Test and align all nozzle manifolds, clearing any blockages to optimize spray performance
- Check all tubing for leaks, confirming secure attachment to manifolds to prevent pressure loss
- Verify all control panel connections are secure, with no frayed or torn wiring that could impact safety
- Confirm fan operation to maintain appropriate internal cooling, replacing if performance is inadequate
- Test evaporative cooling functionality, ensuring performance below ambient temperature
- Test heating bath accuracy, confirming precise temperature alignment with set points
- Verify PSI stability, confirming that pressure is steady and adjustable as needed
- Ensure cosmetic condition is within standards, free from damage impacting function or appearance
Technician Inspection and Recertification Checklist: Blowdown Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Verify internal power supply output voltage
|
Completed
|
|
Confirm fuse holder is secured
|
Completed
|
|
Inspect glass window for cracks or chips
|
Completed
|
|
Assess power-up diagnostics, display, and push wheel buttons
|
Completed
|
|
Ensure pressure gauge and regulator functionality
|
Completed
|
|
Test and align all nozzle manifolds, clear blockages
|
Completed
|
|
Check tubing for leaks and secure attachment to manifolds
|
Completed
|
|
Verify secure control panel connections and check wiring
|
Completed
|
|
Confirm fan operation, replace if needed
|
Completed
|
|
Test evaporative cooling below ambient temperature
|
Completed
|
|
Confirm heating bath temperature accuracy
|
Completed
|
|
Verify PSI stability and adjustability
|
Completed
|
|
Ensure cosmetic condition meets standards
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|