Waters In-Line Degasser AF


The Waters In-Line Degasser AF provides HPLC systems with an automatic, continuous method for removing dissolved gasses from mobile phases. The continuous running pump cycles between high-speed, to provide rapid vacuum pull-down, and low-speed, to sustain a consistent vacuum level. This unique approach enables rapid vacuum equilibration for faster eluent equilibration. The In-Line Degasser AF's continuous operating mode improves chromatographic performance for high-sensitivity applications by eliminating detector baseline drift caused by pump hysteresis in discontinuous operating modes.


All of our equipment is fully tested to perform at factory specifications before being shipped.
INCLUDED:
The Waters In-Line Degasser AF includes:
  • > Power Cord
  • > Operator's Manual
  • > Full 180 Day Parts and Labor Warranty

Waters In-Line Degasser AF Specifications

Chemical Resistance

Unaffected by full range of organic solvents and aqueous solutions of acids, bases, salts, and surfactants

pH Range

0 to 14

Gas Removal

Varies with flow rate

Typical Operating Flow Range

200l/min to 5ml/min

Equilibration time

<1 hour

Pressure Drop

<0.08 psi (0.55 kPa) at 1 mL/min., Milli-Q

water, STP

Wetted Surfaces

PPSa, Teflon AFb, Tefzelc

Inputs

Enable, Disable (External control)

Outputs

dc voltage (Relative vacuum)

Eluent Connections

1/4-28 reversed ferrule fittings;

4 inlet and 4 outlet fittings

Vacuum Chambers

One chamber/eluent, two channels standard, third or fourth channel optional;

<0.5 mL internal volume/channel

Tubular Membrane

1/vacuum chamber, Teflon AF

Vapor Exhaust

1/8-inch (3 mm) barbed fitting

Vacuum Source

Built-in 2-head diaphragm pump, solvent resistant

Vacuum Sensor

Detects vacuum from 0 to 15.6 psiA

Maximum Pressure on Inlet and Outlets

10 psi (70 kPa)

Dimensions

Width: 5.25 inches (13.33 cm)

Height: 5.75 inches (14.60 cm)

Depth: 12.0 inches (30.48 cm)

Power

115 Vac (85 to 132 V), 50/60 Hz

230 Vac (187 to 264 V), 50/60 Hz

Autosensing

Weight

7.5 lbs (3.4 kg)

Ambient Temperature

4 to 40 C (39 to 104 F)

Relative Humidity

10 to 90%, noncondensing