VWR Symphony CO2 Incubator w/ Sterilization Cycle, Cat. #: 98000-362
The VWR Symphony CO2 Incubator w/ Sterilization Cycle, Cat. #: 98000-362 is a state-of-the-art cell culture incubator that combines advanced technology with user-friendly features. This incubator has a spacious 5.3 cubic feet (151L) volume and operates within a temperature range from Ambient +3°C to 55°C, ensuring a stable and uniform temperature environment (± 0.6°C at 37°C for uniformity, and ± 0.1°C at 37°C for stability). It is equipped with a Thermal Conductivity (TC) CO2 sensor, supporting a CO2 range of 0 to 20%. A key feature of this model is its integral 90°C moist heat high-temperature decontamination cycle, which has been proven effective in eliminating a wide range of contaminants. The sterilization cycle efficiently maintains 90°C on all internal surfaces for 25 hours, ensuring thorough decontamination. This process has been validated by independent laboratory tests for its efficacy against various contaminants.
The incubator utilizes advanced air-jacketed temperature control technology, ensuring an optimal and stable environment for cell and tissue cultures. This design not only provides low maintenance but also offers a lighter alternative to traditional water-jacketed models. The unique pan-less humidity water reservoir improves recovery characteristics and cell growth, enhancing the overall incubation process. For user convenience, the incubator features internal shelves and supports that can be easily installed and removed without tools. The sterilization cycle can be initiated with a simple button press. The user interface is designed for ease of use, with large LED displays for temperature and CO2 parameters, touch-sensitive membrane controls, and an intuitive control panel. This provides clear status updates on incubator functions and parameters. Additionally, the incubator includes a fully automated start-up routine, allowing quick and efficient setup. Its intuitive interface and simplified control panel make it easy to use, suitable for a wide range of laboratory settings. All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR Symphony CO2 Incubator w/ Sterilization Cycle, Cat. #: 98000-362 includes:
- > 2 x Shelves
- > Integrated Humidity Reservoir
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR Symphony CO2 Incubator w/ Sterilization Cycle, Cat. #: 98000-362 Specifications
Temperature Range
|
Ambient +3°C to 55°C
|
Temperature Uniformity
|
± 0.6°C at 37°C
|
Temperature Stability
|
± 0.1°C at 37°C
|
CO2 Range
|
0 to 20%
|
CO2 Control Deviation
|
± 0.1%
|
CO2 Accuracy
|
± 0.3 %
|
CO2 Sensor Type
|
TC (Thermal Conductivity)
|
Required Gas Pressure
|
Minimum of 0.8 bar to maximum of 1 bar (12 to 15 PSI)
|
Relative Humidity
|
~ 93% at 37°C
|
Volume
|
5.3 cu. ft. (151L)
|
Sterilization Cycle Temperature
|
90°C on all internal surfaces
|
Sterilization Cycle Duration
|
25 hours
|
Access Port Diameter
|
1.25 in. (32 mm)
|
Data Outputs
|
RS 232
|
Shelf Dimensions (W x D)
|
16.7 x 17.9 in. (42.3 x 45.5 cm)
|
Internal Dimensions (W x D x H)
|
18.5 x 23.9 x 20.9 in. (47 x 60.7 x 53 cm)
|
External Dimensions (W x D x H)
|
25.1 x 34.1 x 30.2 in. (63.7 x 86.7 x 76.6 cm)
|
Weight
|
154.3 lb. (70 kg.)
|
Electrical Requirements
|
120V, 50/60 Hz, 5.2A, 620W
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|