Popular Alternative Choice
Benchmark BT3001 Orbi-Shaker Orbital Shaker
- Powerful shaker with 14 lb capacity at 300 RPM.
- Non-slip platform: 13 x 12 in; Suitable for flasks & dishes.
- Optional MAGic Clamp Platform and Flask Clamps for easy flask size interchangeability.
- Dimensions: 11 x 13 x 3 in; Electrical: 120V, 600W.
Go to Product Details
No Lead Time and Free Shipping!
VWR Orbital Shaker Model 1000
The VWR Orbital Shaker Model 1000, Cat # 89032-088, is a space-saving, low-profile variable speed orbital shaker perfect for use in fume hoods, cold rooms, and incubators. A basic speed knob provides control for consistent uniform mixing action from 40 to 300 RPM. This shaker features gentle ramping to setpoint speed to prevent sample loss. Spill-resistant design protects shaker from damage in case of spills. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The VWR Orbital Shaker Model 1000 includes:
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
VWR Orbital Shaker Model 1000 Specifications
Speed Range
|
40 to 300 RPM
|
Orbit
|
15mm
|
Tray Dimensions (L x W)
|
22.2 x 29.8 cm (8.75 x 11.75 in)
|
Maximum Load
|
3.6 kg (8 lbs)
|
Overall Dimensions (W x D x H)
|
28 x 43.2 x 10.8 cm (11 x 17 x 4.25 in)
|
Electrical Requirements
|
120V, 50/60Hz, 50W
|
VWR 1000 Shaker ManualClick Here to View the Technical Specs for the VWR 1000
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|