VWR Gravity Convection Incubator, Cat # 414005-128
The VWR Gravity Convection Incubator is for basic incubation with reduced air changes. With a temperature range of an ambient +8°C to 65°C it is accurate within ± 0.3°C at 37°C and a heating time of 25 min. to 37°C and 45 min. to 50°C. It provides a safe environment while reducing drying samples out. The temperature sensor allows for repeatability and stability over an extended period of time. It has a digital backlit LCD display for easy viewing. All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR Gravity Convection Incubator, Cat # 414005-128 includes:
- > 2 Shelves
- > Power Cord
- > Operator’s Guide
- > Full 180 Day Parts and Labor Warranty
VWR Gravity Convection Incubator, Cat # 414005-128 Specifications
Capacity
|
32L (1.1 cu ft)
|
Internal Dimensions (L x W x H)
|
11.4 x 12.2 x 14.2 inches
|
External Dimensions (L x W x H)
|
19.1 x 18.9 x 24.6inches
|
Heater
|
150 W
|
Temperature Range
|
Ambient Temperature +8°C to 65°C
|
Temperature Fluctuation
|
± 0.3°C at 37°C / 50°C
|
Temperature Variation
|
± 0.7°C at 37°C, ± 1.2°C at 50°C
|
Temperature Sensor
|
PT 100
|
Control Sensitivity
|
± 0.1°C
|
Temp Display Sensitivity
|
± 1°C
|
Heating-up Time
|
25 min. to 37°C / 45 min to 50°C
|
Recovery (door open 30 sec)
|
11 min to 37°C / 16 min to 50C
|
Weight
|
70.7 lbs
|
Power Supply
|
120V, 60 Hz
|
Shelves
|
2 Stainless Steel
|
Circulation
|
Gravity Convection Type
|
Ventilation
|
1 ea Ventilation Hole w/SS Cap, Dia 40mm
|
Operator's Guide
Click Here to View all the Technical Specs for the VWR Gravity Convection Incubator
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|