VWR Gravity Convection Incubator, Cat. #: 414005-132
The VWR Gravity Convection Incubator, Cat. #: 414005-132 combines microprocessor-controlled temperature with a state-of-the-art design that delivers a true naturally-occurring gravity convection while offering improved temperature uniformity for a wide range of general-purpose incubation applications. This provides precise temperature control from Ambient +8°C to 65°C with a uniformity of ± 1.2°C at 37°C and 50°C.
Additional features include a 3.3 cm (1.3 in.) stainless steel ventilation cap, two doors (one solid, one glass), high temperature grade foamed silicone rubber door gaskets, over-temperature and over-current protection, sensor error detection, and more. The VWR Gravity Convection Incubator is constructed with a durable, powder-coated steel exterior and stainless-steel interior. A locking mode is featured and assists in preventing unintended temperature changes. Users may also take advantage of the 120V, 60Hz, 1A interior outlet.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The VWR Gravity Convection Incubator, Cat. #: 414005-132 includes:
- > 2 Shelves
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR Gravity Convection Incubator, Cat. #: 414005-132 Specifications
Temperature Range
|
Ambient +8°C to 65°C
|
Temperature Uniformity
|
± 1.2°C at 37°C / 50°C
|
Temperature Stability
|
± 0.3°C at 37°C
|
Volume
|
3.7 cu. ft. (105L)
|
Heat-Up Time
|
45 min. to reach 50°C
|
Recovery Time (Door Open for 30 Sec.)
|
12 - 16 min. to reach 50°C
|
Timer
|
99 hr, 59 min. with delay/continuous function
|
Display
|
Digital LCD
|
Controller
|
Advanced, adaptive microprocessor control with rotary knob
|
Safety Devices
|
Over temperature protector, over current protector, sensor error detector
|
Internal Dimensions (L x W x H)
|
16.1 x 19.1 x 21.1 in. (41 x 48.5 x 53.5 cm)
|
External Dimensions (L x W x H)
|
23.8 x 25.4 x 33.1 in. (60.4 x 64.4 x 84.1 cm)
|
Weight
|
130.1 lb. (59 kg)
|
Electrical Requirements
|
120V, 60Hz, 300W
|
- Verify power supply and PCB outputs are within specified voltage ranges
- Inspect display for any dim or dead pixels, replacing as needed to ensure visibility
- Confirm full functionality and responsiveness of all control buttons
- Examine all internal wiring for frays, cuts, or damage, ensuring secure and reliable connections
- Inspect door seals and gaskets for integrity, replacing any compromised components
- Evaluate door handle and locking mechanism for wear or issues, replacing if performance is impaired
- Inspect and clean internal fan; replace if performance does not meet requirements
- Calibrate set temperature using a NIST-traceable probe for precise accuracy
- Confirm functionality of over-temperature alarm and indicator light
- Verify internal outlet functionality, if applicable, to ensure it meets operational standards
- Inspect internal lighting for full functionality and brightness, if applicable
- Perform a cosmetic inspection to confirm the oven’s exterior and interior meet quality standards
Technician Inspection and Recertification Checklist: Ovens
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Verify power supply and PCB outputs are within specified voltage ranges
|
Completed
|
|
Inspect display for dim or dead pixels; replace if needed
|
Completed
|
|
Confirm full functionality of all control buttons
|
Completed
|
|
Examine internal wiring for damage; ensure secure connections
|
Completed
|
|
Inspect door seals and gaskets, replacing any compromised components
|
Completed
|
|
Evaluate door handle and locking mechanism, replacing if worn
|
Completed
|
|
Inspect and clean internal fan; replace if performance is insufficient
|
Completed
|
|
Calibrate set temperature using a NIST-traceable probe
|
Completed
|
|
Confirm over-temperature alarm and indicator light functionality
|
Completed
|
|
Verify internal outlet functionality, if applicable
|
Completed
|
|
Inspect internal lighting for full functionality and brightness
|
Completed
|
|
Perform cosmetic inspection to confirm quality standards of exterior and interior
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|