VWR 5000 Standard Analog Shaker
Speed Range 25 to 500 RPM
Max. Platform Size 18 x 24 in.
- Speed range of 25 to 500 RPM for versatile applications
- Speed accuracy of ±1% above 100 RPM and ±1 RPM below 100 RPM
- Generous 18 x 24-inch (45.7 x 61 cm) platform for increased capacity
- Analog dial control for precise and simple operation
- Timer range of 1 to 120 minutes for controlled mixing protocols
- 1-inch (25.4 mm) orbit for consistent, reliable shaking
- Compact dimensions of 23.6 x 26.7 x 6.0 inches (59.9 x 67.8 x 15.2 cm) to save bench space
- Operates at 120V, 5 amps, and 75 watts for efficient performance
- Durable construction for demanding lab environments
- Ideal for bacterial suspensions, staining/destaining, and cell culture procedures
The VWR 5000 Standard Analog Shaker provides precise and reliable shaking for applications such as bacterial suspensions, staining/destaining, and cell culture procedures that demand consistent, reproducible results. With a speed range of 25 to 500 RPM and an orbit of 1 inch (25.4 mm), this shaker ensures accurate mixing for a wide range of protocols.
Designed for durability, this orbital shaker features an 18 x 24-inch (45.7 x 61 cm) platform and robust construction, capable of handling demanding lab environments. The analog dial control allows for simple operation, enabling precise adjustments to speed and timer settings, with a timer range of 1 to 120 minutes.
Built to fit seamlessly into your lab, the shaker’s compact dimensions (23.6 x 26.7 x 6.0 inches) make it an efficient addition to limited bench space. Operating on a 120V system with 5 amps and 75 watts, this shaker offers dependable performance for various laboratory workflows. This shaker will be fully tested to perform at factory specifications before being shipped.
The VWR 5000 Standard Analog Shaker includes:
- > Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo MaxQ 2000 Analog Orbital Shaker Specifications
Speed Range
|
25 to 500 RPM
|
Speed Accuracy
|
± 1% of set speed (Above 100 RPM)
± 1 RPM (Below 100 RPM)
|
Max. Platform Size
|
18 x 24 in. (45.7 x 61 cm)
|
Timer
|
1 to 120 minutes
|
Control
|
Analog
|
Orbit
|
1 in. (25.4 mm)
|
Overall Dimensions (L x W x H)
|
23.6 x 26.7 x 6.0 in. (59.9 x 67.8 x 15.2 cm)
|
Electrical Requirements
|
120V, 5 amps, 75 watts
|
- Validate that the power supply and PCB deliver stable, correct voltage outputs across all necessary connections, ensuring reliable operation
- Thoroughly inspect the display for any dim, flickering, or dead pixels; replace affected elements to maintain full readability and operational clarity
- Test the full range of button functionality, confirming responsive, accurate input recognition across all settings
- Conduct a detailed inspection of bearings and bearing housing, identifying any worn or damaged bearings, and replace as needed to prevent operational disruption
- Examine the drive belt for signs of cracking, fraying, or material breakdown; replace as required to ensure sustained performance
- Measure and adjust motor drive belt tension to optimal specifications, securing reliable and efficient power transmission
- Verify accuracy of the onboard timer, confirming proper functionality and replacing if it deviates from expected performance
- Calibrate shaker rotation speed with a NIST-traceable tachometer, validating precision within specified operational limits
- Run the shaker continuously for an 8-12 hour testing period, closely observing the shaking platform for smooth, consistent motion, while monitoring for any error indicators such as knocking, skipping, or unusual vibrations
- Perform a comprehensive cosmetic inspection to confirm the device’s exterior meets aesthetic and quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Shakers
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Validate stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim, flickering, or dead pixels; replace as needed
|
Completed
|
|
Test button functionality for responsiveness and accuracy across all settings
|
Completed
|
|
Inspect bearings and housing for wear; replace as needed to prevent disruption
|
Completed
|
|
Examine drive belt for cracking, fraying, or material breakdown; replace as required
|
Completed
|
|
Measure and adjust motor drive belt tension to optimal specifications
|
Completed
|
|
Verify onboard timer accuracy; replace if it deviates from expected performance
|
Completed
|
|
Calibrate shaker rotation speed using a NIST-traceable tachometer
|
Completed
|
|
Run 8-12 hour continuous test; observe platform for smooth, consistent motion
|
Completed
|
|
Perform comprehensive cosmetic inspection for quality standards compliance
|
Completed
|
|