VWR 1570 Benchtop Orbital Incubator Shaker
The VWR 1570 Benchtop Orbital Incubator Shaker stands out for its compact design featuring a transparent hood with a hydraulic lift system, making it ideal for use in limited spaces. This unit is microprocessor-controlled, allowing for precise adjustments and calibration through an easy-to-use keypad. Users can customize the stroke lengths to their specific needs, enhancing the shaker's versatility. The interior and shelves are constructed from stainless steel, ensuring durability and ease of cleaning. Each unit comes equipped with an LED display for clear, easy monitoring, and includes a rotation platform that is self-centering for straightforward installation. The incubator's design also incorporates a user-adjustable counterbalance, offering optimal load flexibility and accommodation for off-center loads.
The VWR 1570 features a speed range of 30 to 400 RPM and a temperature range of Ambient +5°C to 60°C, with exceptional temperature uniformity and sensitivity of ± 0.5°C and ± 0.1°C at 37°C, respectively. The chamber has a generous volume of 4.0 cubic feet (113 liters), and the platform measures 17.3 x 17.3 inches (44 x 44 cm), providing ample space for various sample sizes. Additionally, the incubator includes visual safety lamps as alarms, ensuring the user is alerted to any necessary adjustments, thereby maintaining the integrity of the incubated samples. This makes the VWR 1570 Benchtop Orbital Incubator Shaker a reliable choice for laboratory professionals seeking precision, flexibility, and efficiency in their sample incubation processes.
All of our equipment will be fully tested to perform at factory specifications before being shipped.
The VWR 1570 Benchtop Orbital Incubator Shaker includes:
- > 17.3 x 17.3 in. Platform
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
VWR 1570 Benchtop Orbital Incubator Shaker Specifications
Speed Range
|
30 to 400 RPM
|
Temperature Range
|
Ambient +5°C to 60°C
|
Temperature Uniformity
|
± 0.5°C at 37°C
|
Temperature Sensitivity
|
± 0.1°C at 37°C
|
Chamber Volume
|
4.0 cu. ft. (113L)
|
Orbit
|
User adjustable stroke lengths of 0.5", 0.75", or 1"
|
Alarms
|
Visual safety lamps
|
Platform Size
|
17.3 x 17.3 in. (44 x 44 cm)
|
Chamber Dimensions (W x D x H)
|
17.3 x 17.3 x 17.3 in. (44 x 44 x 44 cm)
|
Overall Dimensions (W x D x H)
|
29.1 x 27.2 x 40.9 in. (74 x 69 x 104 cm)
|
Weight
|
117 lb. (53 kg)
|
Electrical Requirements
|
120V, 50/60Hz, 1100W
|
- Examine the chamber glass for cracks or chips, replacing it as needed to maintain structural integrity.
- Inspect door seals for signs of wear or damage, replacing them to ensure a tight seal.
- Check the display for brightness and clarity, replacing any dim or dead pixels to maintain optimal visibility.
- Test all buttons to confirm full functionality and consistent responsiveness.
- Replace all platform assembly bearings to ensure smooth and reliable motion.
- Inspect the drive belt for wear, replacing it if torn or cracked to maintain performance.
- Examine all wiring for frays, tears, or breakages, addressing any issues for electrical safety.
- Verify each PCB test point, replacing any malfunctioning components to ensure system reliability.
- Check drive belt tension to confirm correct alignment and function.
- Confirm the compressor’s operation, cleaning condenser filter and fins for efficient cooling.
- Calibrate internal temperature accuracy using an NIST-traceable temperature probe.
- Measure and calibrate platform RPMs with an NIST-traceable tachometer to ensure precision.
- Conduct a final cosmetic inspection to verify all surfaces meet appearance standards.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Examine chamber glass for cracks or chips, replace as needed | Completed | |
Inspect door seals for wear, replace to ensure tight seal | Completed | |
Check display brightness, replace dim or dead pixels | Completed | |
Test all buttons for full functionality and responsiveness | Completed | |
Replace platform assembly bearings for reliable motion | Completed | |
Inspect drive belt for wear, replace if torn or cracked | Completed | |
Examine wiring for frays or breakages, repair as needed | Completed | |
Verify PCB test points, replace faulty components | Completed | |
Check drive belt tension for proper alignment | Completed | |
Confirm compressor functionality, clean filter and fins | Completed | |
Calibrate internal temperature with NIST probe | Completed | |
Calibrate platform RPM with NIST-traceable tachometer | Completed | |
Conduct cosmetic inspection for appearance standards | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|