VWR 1535 Digital Incubator
Temperature Range Ambient +5°C to 70°C
Temperature Uniformity ± 0.25°C at 37°C
- Quick temperature recovery ensures efficient operation
- Full-length inner glass door allows specimen viewing without atmospheric disturbance
- Microprocessor control for easy parameter setting and calibration
- Temperature display can be calibrated to match a secondary thermometer
- Provides temperature stability of ± 0.1°C
- Includes over-temperature protection with a secondary independent safety controller
- Visual alarm activates when safety limits are exceeded
- Through-wall access port accommodates a chart recorder
The VWR 1535 Digital warm air incubator provides quick temperature recovery and has a full-length inner glass door for viewing specimens without disturbing the chamber atmosphere. Microprocessor control allows easy keypad parameter setting and calibration. The temperature display may also be calibrated to match a secondary thermometer. For over-temperature protection, a second, independent safety controller can be preset and has visual alarm when activated. Features through-wall access port for chart recorder. The VWR 1535 is a 6.7 cubic foot unit that offers temperatures from 5°C above ambient to 70°C with a uniformity of ± 0.25°C and stability of ± 0.1°C. All of our equipment is fully tested to perform at factory specifications before being shipped.
The VWR 1535 Digital Incubator includes:
- > Calibration
- > 2 Shelves
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
VWR 1535 Digital Incubator Specifications
Temperature Range
|
Ambient +5°C to 70°C
|
Temperature Uniformity
|
± 0.25°C @ 37°C
|
Temperature Stability
|
± 0.1C
|
Capacity
|
6.7 cu ft
|
Interior Dimensions (W x D x H)
|
24 x 24 x20 in
|
Exterior Dimensions (W x D x H)
|
30 x 30 x32 in
|
Electrical Requirements
|
115V, 50/60Hz, 5A
|
VWR 1535 Digital Incubator Manual
Click Here to View all the Technical Specs for the VWR 1535
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|