Popular Alternative Choice
Benchmark H2200-H MyTemp Mini Digital Incubator
- MyTemp Mini Incubator: Digital temperature control with real-time LED display.
- Design: Compact 14.5" x 13" footprint, houses up to 2L flasks, two adjustable shelves, and an internal outlet.
- Specifications: 20L capacity, heats from Ambient +5C to 75C, 10.3 x 9.3 x 12.8in interior dimensions.
- Included: Two Shelves, Internal Outlet, Power Cord, Guide, Manufacturer's Docs, 2 Year Warranty.
Go to Product Details
Brand New!
VWR 1500E Incubator
Capacity 1 cu.ft. (28.3 L)
Temperature Range Ambient +5C to 70C
Temperature Uniformity ± 0.5C at 37C
- Designed for routine small incubation loads utilizing gravity convection.
- Efficient heating is supplied by economical, low-watt density strip heaters.
- Temperature is regulated by a hydraulic temperature controller.
- Galvanized, cold-rolled steel chamber is insulated with 5.1cm of industrial fiberglass to minimize heat loss and maximize uniformity.
- Glossy white interior allows for easy contamination detection.
The VWR 1500E Incubator is a 1 cubic foot incubator with a temperature range of 5°C above ambient to 70°C. Temperature uniformity is ± 0.5°C at 37°C. The VWR 1500E series was designed for routine small incubation loads utilizing gravity convection. The stainless steel chamber design minimizes heat loss while maximizing uniformity. All of our equipment will be fully tested to perform at factory specifications before being shipped.
The VWR 1500E Incubator includes:
- > Operators Manual
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
VWR 1500E Incubator Specifications
Temperature Range
|
Ambient +5°C to 70°C
|
Temperature Uniformity
|
± 0.5°C at 37°C
|
Number of Shelves
|
2
|
Interior Dimensions (W x H x D)
|
12 x 14 x 12 in
|
Exterior Dimensions (W x H x D)
|
16 x 22 x 16 in
|
Electrical Requirements
|
120V, 50/60Hz, 3.0A
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|