Thermo Scientific Thermal Mixer w/ 1.5ml Block, Cat. #: 13687711
The Thermo Scientific Thermal Mixer w/ 1.5ml Block, Cat. #: 13687711 provides mixing and temperature control of samples in micro test tubes or PCR plates with corresponding blocks. The features of this thermal mixer meet the highest expectations of users according to many parameters, including fast reaching of specified mixing speeds from 250 to 1,400 RPM and stability of maintaining the set temperature in a range from Ambient -15°C to 100°C through the block surface.
With the help of the temperature calibration function, the user can calibrate the unit approximately ± 6% of the selected temperature to compensate for any differences in the thermal behavior of tubes from varying manufacturers. An LCD display is also featured, which indicates pre-set and current values of temperature, speed, and time of operation. Functions of heating and mixing can be performed either simultaneously or independently as well.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Thermal Mixer w/ 1.5ml Block, Cat. #: 13687711 includes:
- > 24 x 1.5ml Block
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Thermal Mixer w/ 1.5ml Block, Cat. #: 13687711 Specifications
Speed Range
|
250 to 1,400 RPM
|
Speed Increments
|
10 RPM
|
Maximum Speed Deviation
|
2% at 250 RPM or 0.7% at 1400 RPM
|
Temperature Range
|
Ambient -15°C to 100°C (Minimum Setpoint of 4°C)
|
Temperature Uniformity
|
± 0.1°C at 37°C
|
Temperature Stability
|
± 0.1°C at 37°C
|
Temperature Increments
|
± 0.1°C
|
Heating/Cooling Rate
|
Heating: 5°C/minute from 25°C to 100°C
Cooling: 5°C/minute from 100°C to 25°C or 1.8°C from 25°C to 4°C
|
Orbit
|
2 mm
|
Timer
|
1 minute to 96 hours, or continous
|
Display
|
16 x 2 symbol LCD screen
|
Dimensions (W x D x H)
|
5.11 x 9.05 x 8.07 in. (13 x 23 x 20.5 cm)
|
Weight
|
8.82 lb. (4 kg)
|
Electrical Requirements
|
12VDC via 120/240 AC Adapter, 6A, 60W
|
- Confirm stability and accuracy of power supply and PCB voltage outputs
- Inspect display integrity, ensuring pixel clarity and replacing any degraded or non-functional elements
- Verify complete functionality and responsiveness of all control buttons, switches, and knobs
- Conduct thorough inspection of internal wiring and all electrical connections for continuity; replace or secure any degraded or substandard components
- Examine block holder platform for alignment, obstructions, or error-inducing irregularities
- Perform detailed inspection of motor bearings, assessing for wear or misalignment, and replace if performance deviates from specifications
- Calibrate rotational speed precisely using a NIST-traceable tachometer for compliance with operational standards
- Calibrate temperature accuracy using a NIST-traceable temperature probe, validating readings across the operational range
- Ensure timer precision by validating functionality and consistency against set parameters
- Conduct a comprehensive cosmetic inspection, verifying all surfaces and external components meet quality standards
Technician Inspection and Recertification Checklist: Thermo Mixer
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stability and accuracy of power supply and PCB voltage outputs
|
Completed
|
|
Inspect display integrity, ensuring pixel clarity and replacing any degraded elements
|
Completed
|
|
Verify functionality and responsiveness of all control buttons, switches, and knobs
|
Completed
|
|
Inspect internal wiring and connections for continuity, replacing or securing degraded components
|
Completed
|
|
Examine block holder platform for alignment, obstructions, or irregularities
|
Completed
|
|
Inspect motor bearings for wear or misalignment, replacing if necessary
|
Completed
|
|
Calibrate rotational speed using an NIST-traceable tachometer
|
Completed
|
|
Calibrate temperature with an NIST-traceable probe, validating accuracy
|
Completed
|
|
Ensure timer precision by validating against set parameters
|
Completed
|
|
Conduct cosmetic inspection for quality standard compliance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|