The Thermo Scientific TSX3020FD High-Performance -20C Lab Freezer, 230V is a state-of-the-art solution, meticulously designed for the uncompromised storage of sensitive materials such as enzymes, reagents, pharmaceuticals, vaccines, and various biological substances commonly utilized in laboratory and clinical settings. One of its standout features is the innovative V-drive technology, which intelligently adapts compressor speed to usage patterns, efficiently balancing energy savings with rigorous protection standards. This unique system ensures optimal temperature uniformity, particularly during periods of high demand such as door openings, by accelerating compressor activity, and conserves energy during stable periods by reducing speed. Enhanced with cold wall convection cooling, the TSX3020FD maintains stable and uniform temperatures throughout, essential for critical storage requirements. Its user-friendly microprocessor control enables precise temperature management, augmented by an interior light, setpoint security, and integral alarms to safeguard contents reliably. For additional security, standard door locks are installed, delivering peace of mind. Beyond security and performance, the TSX3020FD is designed with the user’s physical comfort and convenience in mind. The ergonomic design features include a door handle with a soft-touch grip, an angled controller to prevent neck strain, and an automatic pressure equalization port for quick reentry following door closures. The unit also boasts easy-to-clean, adjustable epoxy-coated shelves, contributing to the overall maintainability of the freezer. Furthermore, the TSX3020FD respects the need for a quiet work environment, operating at a whisper-quiet 52 dBA to ensure laboratory staff can work undisturbed. The practicality is further enhanced by a self-closing door mechanism, featuring a 90° stop that facilitates convenient inventory management and loading procedures. Overall, the TSX3020FD combines advanced technology with user-centric design, delivering a high-performance freezing solution that does not compromise on security, energy efficiency, or operational comfort. All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific TSX3020FD High-Performance -20C Lab Freezer, 230V includes:
- > Key
- > 4 x
Shelves
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific TSX3020FD High-Performance -20C Lab Freezer, 230V Specifications
Temperature Range
|
-25°C to -15°C (-20°C factory preset)
|
Volume
|
29.2 cu. ft. (827L)
|
Refrigerant
|
R290
|
Defrost Method
|
Manual
|
Monitoring Option
|
Wireless
|
Internal Dimensions (D x W x H)
|
28.5 x 30.0 x 58.0 in. (72.3 x 76.2 x 147.3 cm)
|
External Dimensions (D x W x H)
|
38.9 x 34.0 x 78.5 in. (99 x 86.4 x 199.4 cm)
|
Weight
|
483 lb. (219 kg.)
|
Electrical Requirements
|
208-230V, 60Hz, 15A
|
- Inspect the display for pixel clarity and brightness, replacing any dim or dead pixels.
- Test all buttons for responsiveness, resistance, and durability; replace any that don’t meet standards.
- Check the door seal and lock for airtightness, replacing any compromised components.
- Run diagnostics on PCB test points for voltage and current accuracy, replacing faulty components.
- Verify compressor output and airflow, ensuring all internal fans function properly.
- Cross-check internal temperature probe accuracy with a NIST-traceable probe, recalibrating if needed.
- Inspect wiring for frays or tears, replacing compromised sections as necessary.
- Test the start capacitor's capacitance and replace if it’s outside the specified range.
- Replace all internal filters and thoroughly clean condenser fins for efficient heat dissipation.
- Verify the alarm battery holds a full charge and replace if there is any degradation.
- Operate the unit for 72 hours to confirm stable temperatures and operational consistency.
- Perform a final cosmetic check, repairing or touching up surfaces as needed for a professional appearance.
Technician Inspection and Recertification Checklist
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task |
Completed? |
Remarks? |
Inspect display for pixel issues and replace if needed | Completed | |
Test button functionality and replace any defective buttons | Completed | |
Examine door seal and lock functionality; replace if compromised | Completed | |
Conduct PCB test point diagnostics and replace as needed | Completed | |
Verify compressor and internal fan functionality | Completed | |
Confirm temperature probe accuracy with NIST-traceable probe | Completed | |
Inspect all wiring for frays, tears, or breakages | Completed | |
Test start capacitor capacitance range; replace if needed | Completed | |
Replace all internal filters and clean condenser fins | Completed | |
Test alarm battery for full functionality | Completed | |
Run 72-hour test with data recorder to confirm stable temperatures | Completed | |
Conduct cosmetic inspection and address any issues | Completed | |
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|