Thermo Scientific Steri Cycle 370 CO2 Incubator Dual Stack
The Thermo Scientific Steri Cycle 370 CO2 Incubator Dual Stack delivers unmatched contamination control and sustained reliability. With a temperature range of Ambient +5°C to 50°C and precise microprocessor PID control ensuring temperature uniformity within ± 0.3°C at 37°C, this incubator maintains optimal conditions for sensitive cultures. It features a CO2 range of 1 to 20%, controlled to ± 0.1%, and utilizes a heat-resistant TC sensor for accurate monitoring. The system boasts a robust direct-heat design with an in-chamber HEPA air filtration system that continually filters out particulates, maintaining Class 100 air quality.
The Steri Cycle 370 includes a high-temperature sterilization cycle at 140°C across all internal surfaces for approximately 12 hours, offering an efficient solution for routine decontamination. Additional features include an intuitive alphanumeric display for monitoring temperature and CO2 levels, configurable alarms for security and maintenance reminders, and a polished stainless-steel interior with removable shelves for easy cleaning. This dual-stack incubator ensures operational efficiency and high-performance standards, making it an essential tool for advanced research environments.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Steri Cycle 370 CO2 Incubator Dual Stack includes:
- > 3 x Shelves Per Unit
- > 2 x Humidity Pans
- > Stacking Kit
- > NIST Calibration and Preventative Maintenance Report
- > 2 x Power Cords
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Steri Cycle 370 CO2 Incubator Dual Stack Specifications
Temperature Range
|
Ambient +5°C to 50°C
|
Temperature Uniformity
|
± 0.3°C at 37°C
|
Temperature Control
|
± 0.1°C via microprocessor PID control
|
CO2 Range
|
1 to 20%
|
CO2 Control
|
± 0.1% via microprocessor PID control
|
CO2 Sensor Type
|
TC
|
Required Gas Pressure
|
15 psig (1 bar), ± 5 psig (0.3 bar)
|
Relative Humidity
|
95% at 37°C
|
Volume
|
6.5 cu. ft. (184L) [Per Unit]
|
Sterilization Cycle Temperature
|
140°C on all internal surfaces
|
Sterilization Cycle Duration
|
Approximately 12 hours
|
Access Port Diameter
|
1.25 in. (32 mm)
|
Shelf Dimensions (W x D)
|
18.5 x 18.5 in. (47 x 47 cm)
|
Internal Dimensions (W x H x D)
|
21.3 x 26.8 x 20 in. (54 x 68.1 x 50.8 cm) [Per Unit]
|
External Dimensions (W x H x D)
|
26.1 x 39.5 x 25 in. (66.3 x 100.3 x 63.5 cm) [Per Unit]
|
Weight
|
260 lb. (118 kg.) [Per Unit]
|
Electrical Requirements
|
115V, 50/60 Hz, 9.6A [Per Unit]
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|