Thermo Scientific Steri-Cult 3307 CO2 Incubator w/ Sterilization Cycle
Temperature Range Ambient +5°C to 50°C
Volume 8.2 cu. ft. (232.2L)
Temperature Uniformity ± 0.2°C at 37°C
- Stackable design with polished stainless-steel interior and coved corners for easy cleaning
- Direct heat chamber with an active airflow system to reduce cross-contamination
- Advanced IR CO2 sensor with a range of 0 to 20% and control accuracy of ±0.1% at 5.0% CO2
- Active humidity control system eliminates the need for a water pan and maintains accuracy of ±2.0%
- 140°C sterilization cycle runs for 14 hours to effectively eliminate bacteria, mold, yeast, and mycoplasmas
- ISO Class 5 HEPA filtration system filters the chamber volume every 60 seconds
- Enviro-Scan Microprocessor Control with alphanumeric LED display and message center for alarm notifications
- Removable stainless-steel shelves measuring 17.7 x 19.9 in for easy cleaning and flexibility
- External water reservoir for convenient and contamination-free humidity management
- Built-in over-temperature protection fixed at 160°C for enhanced safety
The Thermo Scientific Steri-Cult 3307 CO2 Incubator w/ Sterilization Cycle is a high-capacity, stackable incubator designed for extensive culturing while ensuring superior contamination control. It features a direct heat chamber optimized for high throughput applications, with a polished stainless-steel interior and coved corners for easy cleaning. This incubator provides for a temperature range of Ambient +5°C to 50°C with a uniformity of ± 0.2°C at 37°C. The chamber's active airflow system ensures homogeneous conditions and reduces cross-contamination risks, supported by sturdy, easily removable stainless-steel shelves. The Steri-Cult 3307 boasts an active humidity control system with a range from Ambient to 90% and accuracy of ± 2.0%, eliminating the need for a water pan and featuring an external water reservoir to reduce chamber contamination. Additionally, the incubator has a 140°C sterilization cycle, which operates overnight for 14 hours to eliminate bacteria, mold, yeast, and mycoplasmas, thereby simplifying and automating cleaning. The Steri-Cult features a CO2 range from 0 to 20% and utilizes an advanced IR CO2 sensor, providing precise monitoring unaffected by humidity changes The ISO Class 5 HEPA filtration airflow system continuously filters the entire chamber volume every 60 seconds, reaching ISO Class 5 air quality within five minutes after a door opening. This feature minimizes contaminants from settling on interior surfaces. Operational control is managed via the Enviro-Scan Microprocessor Control, which includes an easy-to-use alphanumeric LED display, flexible controls, and a message center for process and alarm status messages.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Steri-Cult 3307 CO2 Incubator w/ Sterilization Cycle includes:
- > 3 x Shelves
- > Diffuser Pan
- > Humidity Bottle
- > New HEPA Filter
- > NIST Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Steri-Cult 3307 CO2 Incubator w/ Sterilization Cycle Specifications
Temperature Range
|
Ambient +5°C to 50°C
|
Temperature Uniformity
|
± 0.2°C at 37°C
|
Temperature Control
|
± 0.1°C at 37°C
|
CO2 Range
|
0 to 20%
|
CO2 Control
|
± 0.1% at 5.0%
|
CO2 Sensor Type
|
IR (Infrared)
|
Inlet Pressure
|
15 PSIG, ± 5 PSIG
|
Relative Humidity Range
|
Ambient to 90%
|
Relative Humidity Control
|
± 2.0%
|
Volume
|
8.2 cu. ft. (232.2L)
|
Sterilization Cycle Temperature
|
140°C
|
Sterilization Cycle Duration
|
Approximately 14 hours
|
Access Port Diameter
|
1.4 in. (3.6 cm)
|
Over Temperature Protection
|
Fixed at 160°C, ± 5.6°C
|
Alarm Contacts
|
Deviation of temperature, CO2, RH, and power failure
|
Shelf Dimensions (W x D)
|
17.7 x 19.9 in. (45.2 x 50.5 cm)
|
Internal Dimensions (W x H x D)
|
20.8 x 32.8 x 20.6 in. (52.8 x 83.8 x 52.3 cm)
|
External Dimensions (W x H x D)
|
35 x 39.4 x 27 in. (88.9 x 100.1 x 68.6 cm)
|
Weight
|
330 lb. (149.7 kg.)
|
Electrical Requirements
|
115V, 50/60 Hz, 10.5A, 1Ph
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|