Thermo Scientific Sorvall RC 6 Plus Refrigerated Centrifuge
Temperature Range 2°C to 40°C
Maximum Capacity 4 x 1000ml
- High-speed refrigerated floor centrifuge for efficient substance separation
- Features a high-frequency brushless motor for rapid separations and reduced maintenance requirements
- Precise gravitational force settings and display ensure accurate separations with reproducibility
- Includes a pre-cool function to efficiently cool rotors before use, enhancing sample integrity
- Supports a diverse range of rotor types, from lightweight composite to robust aluminum
- Equipped with safety interlocks to prevent operation with an open chamber door or during malfunctions
- Uses environmentally-friendly, CFC-free refrigerant to maintain controlled temperatures
- Touch keyboard and bright LCD for easy monitoring and customizable settings with 9 acceleration and 10 deceleration rates
- Offers 30 program memory slots and step-run functionality for versatile application handling
The Thermo Scientific Sorvall RC 6 Plus Refrigerated Centrifuge is a high-speed refrigerated floor centrifuge that achieves speeds up to 22,000 RPM, facilitating the separation of substances with different densities at controlled temperatures. It is equipped to process up to 4 liters of sample per run and features a high-frequency brushless motor that ensures rapid separations with reduced maintenance. Users can precisely set and display the gravitational force (g-force) for accurate separations and rely on the w2dt integrator to achieve reproducible results across different runs. The instrument includes a pre-cool function for efficient rotor cooling and supports a wide range of rotor types, from lightweight composite to robust aluminum, accommodating various sample sizes. With safety as a priority, the centrifuge has system interlocks to prevent operation if the chamber door is open or to halt a run if a malfunction is detected.
The refrigeration system uses an environmentally-friendly, CFC-free refrigerant, maintaining the desired temperatures effectively during operation. The design offers ease of use with a touch keyboard, bright LCD for at-a-glance monitoring, and customizable settings such as 9 acceleration and 10 deceleration rates. With 30 program memory slots, step-run functionality for flexible application, and real-time control for delayed starts or stops, the centrifuge optimizes workflow. Additional features include automatic RCF/RPM conversion and a user lockout function for secure access in shared lab environments. It is compatible with a range of labware, from small tubes to large bottles, ensuring seamless integration and long-term reliability for various research applications. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Thermo Scientific Sorvall RC 6 Plus Refrigerated Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Sorvall RC 6 Plus Refrigerated Centrifuge Specifications
Maximum Speed
|
22,000 RPM
|
Minimum Speed
|
300 RPM
|
Maximum RCF
|
55,200 x g
|
Maximum Capacity
|
4 x 1000ml
|
Temperature Range
|
2°C to 40°C
|
Temperature Accuracy
|
± 2°C
|
Timer
|
1 minute to 99 hours 59 minutes with continuous run (hold)
|
Acceleration Rates
|
9
|
Deceleration Rates
|
10
|
Display
|
Digital LCD
|
Program Storage
|
Up to 30 programs
|
Noise Level
|
< 64 dB(A)
|
Average Heat Dissipation
|
3.7 kW/h
|
Dimensions (H x W x D)
|
44.5 x 29.6 x 32.8 in. (113.2 x 75.2 x 83.5 cm)
|
Weight
|
770 lb. (350 kg)
|
Electrical Requirements
|
208V, 50/60Hz, 30A, 4900W
|
Click Here to Expand Rotor Specifications
SS-34 Rotor Specs
|
Capacity
|
8 x 50 mL
|
Max RPM
|
20,000
|
Max RCF
|
47,808 x g
|
SLA-1500 Rotor Specs
|
Capacity
|
6 x 250 mL
|
Max RPM
|
14,500
|
Max RCF
|
31,916 x g
|
SLA-3000 Rotor Specs
|
Capacity
|
6 x 500 mL
|
Max RPM
|
11,000
|
Max RCF
|
20,450 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|