Thermo Scientific Sorvall Legend XF Centrifuge
- Maximum speed of 15,200 RPM with a maximum RCF of 25,314 x g for high-performance centrifugation
- Large capacity of 4 x 750mL (3 Liters) for high-throughput applications
- Quiet operation with noise levels below 61 dBa, suitable for all lab environments
- Onboard LCD display for easy reading of parameters and data entry
- Extended timer options with a maximum run time of 99 hours, 59 minutes or indefinite operation
- Program storage capability for up to 99 unique centrifugation programs
- Runs on 120V 60Hz power supply, adaptable to standard lab setups
- Compatible with various adapters depending on the chosen rotor
The Thermo Sorvall Legend XF Centrifuge
features a maximum speed of 15,200 RPM (fixed Angle rotor) and a maximum relative
centrifugal force of 25,314 times gravity. An onboard LCD display makes reading
information easy and entering parameters a breeze. The XF has a maximum timer
run time of 99 hours and 59 minutes and also offers the ability to run indefinitely.
With a maximum capacity of 4 x 1000 milliliters, the Sorvall Legend XF offers
an extremely high throughput. With
advanced sound-reducing technology, the XF features a quiet running
noise level, making it the perfect centrifuge for both smaller and larger labs.
We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.
Please confirm the selected rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our
Rotor Compatibility Guide.
The Thermo Scientific Sorvall Legend XF Centrifuge includes:
- > Rotor Selected Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Sorvall Legend XF Centrifuge Specifications
Maximum Speed
|
15,200 Rotations per Minute
|
Max Capacity
|
4 x 750mL (3 Liters)
|
Maximum RCF
|
25,314 x G
|
Noise level
|
<61dBa
|
Timer
|
99 hours, 59 minutes
|
Program Storage
|
99 Programs
|
Electrical Requirements
|
120V 60Hz
|
Refrigeration |
No Refrigeration |
Click Here to Expand Rotor Specifications
TX-750 Rotor Specs
|
Capacity
|
4 x 750 mL bottles
|
Max RPM
|
4,500
|
Max RCF
|
4,415 x g
|
Fiberlite F-15-8x50Cy Rotor Specs
|
Capacity
|
8 x 50 mL Conical Tubes
|
Max RPM
|
14,500
|
Max RCF
|
24,446 x g
|
Fiberlite F-15-6x100Y Rotor Specs
|
Capacity
|
6 x 100 mL Tubes
|
Max RPM
|
15,000
|
Max RCF
|
24,652 x g
|
M-20 Microplate Rotor Specs
|
Capacity
|
2 x 3 Standard or 2 x 1 Deepwell Plates
|
Max RPM
|
4,000
|
Max RCF
|
2,272 x g
|
Bioliner Rotor Specs
|
Capacity
|
4 x 750 mL bottles
|
Max RPM
|
3,500
|
Max RCF
|
2,739 x g
|
TX-1000 Rotor Specs
|
Capacity
|
4 x 1000 mL Bottles
|
Max RPM
|
3,800
|
Max RCF
|
3,374 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|