Thermo Scientific Savant DNA 130 SpeedVac Vacuum Concentrator
The Thermo Scientific Savant DNA 130 SpeedVac Vacuum Concentrator is a compact vacuum concentrator specifically designed for DNA and RNA applications. This integrated benchtop unit is notable for its small footprint, conserving valuable lab space. It features a corrosion-resistant chamber coated with PTFE, ensuring longevity and ease of maintenance. The built-in, oil-free vacuum pump provides hassle-free operation without the need for regular maintenance. This model is user-friendly, equipped with a straightforward interface and three programmable settings that can be customized and saved for repeated use. Temperature control is precise, ranging from 35°C to 65°C in 5°C increments, and the heat time can be set from 1 minute to 9 hours and 59 minutes, or run continuously. An audible alarm notifies users when a cycle is complete, enhancing convenience and efficiency. Designed with sample protection in mind, the DNA 130 allows for drying at ambient temperatures or with low heat to prevent sample denaturing. The oil-free diaphragm pump ensures a smooth transition to the ultimate vacuum setting, minimizing solvent bumping. Safety is a priority with a PTFE-coated chamber that resists chemicals and is easy to clean, a latched lid with a lock indicator for user safety, and a lid stay mechanism for hands-free operation. The unit is virtually maintenance-free and includes an external post trap for neutralizing ammonium hydroxide vapors. Optional chemical traps and cartridges are available for handling hazardous fumes and chemicals. The DNA 130 is versatile, compatible with solvents such as water, PCR buffers, ethanol-water mixtures, and ammonium hydroxide, though an ammonia post-trap and neutralizing solution accessory are recommended for the latter. With its consistent performance, the DNA 130 is an ideal choice for laboratories seeking efficient, reliable, and safe sample preparation.
All of our equipment is fully tested to perform at factory specifications before being shipped.
The Thermo Scientific Savant DNA 130 SpeedVac Vacuum Concentrator includes:
- > RD2MP Microplate Rotor
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Savant DNA 130 SpeedVac Vacuum Concentrator Specifications
Temperature Range
|
35°C to 65°C
|
Temperature Increments
|
± 5°C
|
Timer
|
1 min. to 9 hrs, 59 min.
|
Ultimate Vacuum
|
< 13 mbar (10 Torr)
|
Displacement
|
36L/min at 60Hz
|
Cover Material
|
Transparent acrylic
|
Vacuum Chamber Material
|
Teflon coated aluminum casing
|
Rotor Capacity
|
2 x Microplates
|
Dimensions (W x D x H)
|
11.5 x 25 x 12 in. (29 x 64 x 31 cm)
|
Weight
|
78.5 lb. (35.5 kg)
|
Electrical Requirements
|
115V, 60Hz, 10A
|
- Inspect the display panel, replacing any dim or dead pixels to ensure clear visual feedback
- Verify full functionality of all buttons, confirming consistent responsiveness
- Examine the acrylic or glass lid for any cracks or chips, ensuring safety and integrity
- Check the vacuum chamber seal for any tears or cracks that could impact pressure retention
- Test chamber lid functionality, confirming all micro switches report a consistent .04 Ohms
- Verify external connections to the pump or cold trap, ensuring they are properly sealed and leak-free
- Inspect and clean the internal fan and condenser (if applicable) for efficient cooling and airflow
- Examine all internal wiring for frays or breakages, confirming secure and intact connections
- Test motor functionality, replacing motor or bearings if performance is compromised
- Inspect internal pump seals, replacing any worn or damaged seals to maintain vacuum integrity
- Inspect all internal hoses for potential leaks, ensuring tight, clean seals
- Confirm unit is capable of achieving all manufacturer-specified temperature ranges
- Verify the unit can reach specified vacuum levels in accordance with manufacturer standards
- Conduct a 6-8 hour cycle test to confirm consistent operational performance
Technician Inspection and Recertification Checklist: Centrifugal Evaporators
The Following Checklist Is to Be Performed on All Incoming and Outgoing Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display panel for dim or dead pixels
|
Completed
|
|
Confirm functionality of all buttons
|
Completed
|
|
Check acrylic/glass lid for cracks or chips
|
Completed
|
|
Inspect vacuum chamber seal for tears or cracks
|
Completed
|
|
Confirm chamber lid micro switches report .04 Ohms
|
Completed
|
|
Verify pump or cold trap connections are sealed and leak-free
|
Completed
|
|
Inspect and clean internal fan and condenser (if applicable)
|
Completed
|
|
Inspect internal wiring for frays or breakages
|
Completed
|
|
Test motor functionality, replace motor/bearings if needed
|
Completed
|
|
Inspect internal pump seals, replace if needed
|
Completed
|
|
Inspect all internal hoses for leaks, ensure clean seals
|
Completed
|
|
Confirm unit achieves specified temperature ranges
|
Completed
|
|
Verify unit reaches specified vacuum levels
|
Completed
|
|
Conduct a 6-8 hour cycle test for operational performance
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|