Thermo Scientific ST16R Refrigerated Centrifuge
Temperature Range
-10°C to +40C
- Auto-lock rotor system for secure and easy rotor exchange.
- Click-seal bucket sealing system, ensuring safe and leak-proof sample containment.
- Enhanced safety features, including steel housing, rotor chamber, impact-resistant plastic front panel, and electronic imbalance detection.
- SMARTspin technology optimizing acceleration and braking, enhancing safety and separation quality.
- Intuitive controls with one-finger lid closure for user-friendly operation.
- Maintenance-free induction motor for quiet, low-vibration operation, ensuring longevity.
The Thermo Scientific ST16R Refrigerated Centrifuge features auto-lock rotor system, click-seal bucket sealing system, SMARTspin technology, rotor detection, electronic imbalance detection, space-saving design, and outstanding energy efficiency. The ST16R is extremely easy to use with intuitive controls and one finger lid close. Easily pre-set speed, RCF value, running time, temperature, and running profile (acceleration and braking curves) for each run. This refrigerated model offers a temperature range of -10°C to 40°C with speeds from 300 to 15,200 RPM, dependent upon rotor used. We may have different adapters available depending on the rotor chosen with this centrifuge. If the selected rotor includes adapters that will not work for your application, please inquire.The Thermo ST16R is recommended for use with cell culture and blood processing. All of our equipment is in excellent condition and will be fully tested to perform at factory specifications before being shipped.
Please confirm this rotor meets your application needs and request a different rotor, if necessary. Available rotors can be found in our Rotor Compatibility Guide.
The Thermo Scientific ST16R Refrigerated Centrifuge includes:
- > Selected Rotor Above
- > NIST Centrifuge Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Guide
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific ST16R Refrigerated Centrifuge Specifications
Temperature Range
|
-10°C to +40C
|
Maximum Speed
|
15,200 RPM w/ Fixed Angle Rotor
|
Minimum Speed
|
300 RPM
|
Maximum RCF
|
25,830 x g
|
Maximum Kinetic Energy
|
< 62.5 kJ
|
Noise Level at Max
|
< 63 dB(A)
|
Dimensions (H x W x D)
|
360 x 623 x 605 mm
|
Electrical Requirements
|
120 VAC, 60 Hz, 9.5 A, 850 W
|
Click Here to Expand Rotor Specifications
TX-200 Rotor Specs
|
Capacity
|
4 x 180 mL bottles
|
Max RPM
|
5,500
|
Max RCF
|
5,580 x g
|
Fiberlite F-15-6x100Y Rotor Specs
|
Capacity
|
6 x 100 mL Tubes
|
Max RPM
|
13,000
|
Max RCF
|
24,652 x g
|
M-20 Microplate Rotor Specs
|
Capacity
|
2 x 3 Standard or 2 x 1 Deepwell Plates
|
Max RPM
|
4,000
|
Max RCF
|
2,272 x g
|
TX-400 Rotor Specs
|
Capacity
|
4 x 400 mL Bottles
|
Max RPM
|
5,000 RPM
|
Max RCF
|
4,696 x g
|
- Door Latch Assembly Checked for Full Function
- Door Hinges are Checked and Lubed as Needed
- Drive Belt is Checked for Rips and Tears and Replaced if Needed
- Rotors and O-rings are Lubed as Needed
- Lamps and Display are Checked and Replaced if Dim
- Line Cord is Checked for Signs of Fray or Damage and Replaced Accordingly
- Internal Circuitry is Checked for Faulty/Shorted Components and Replaced Accordingly
Thermo Scientific ST16R Centrifuge Manual
Click Here to View all the Technical Specs for the ST16R
- Confirm the power supply output voltages match specified requirements for stable distribution to all internal components
- Measure and verify voltages at PCB test points to ensure accurate power delivery and circuit integrity
- Confirm rotor compatibility with the centrifuge model and check that rotor detection systems recognize the rotor type
- Inspect the exterior for any cosmetic or structural damage that could affect performance or safety
- Check the door mechanism alignment to ensure secure and smooth opening and closing
- Inspect and replace the gas strut with an OEM part to maintain proper door support and safety
- Examine the rotor chamber bowl for dents or damage and replace as needed to ensure safe operation
- Replace and confirm full functionality of all lid lock switches to maintain secure closure
- Inspect and lubricate the door lock mechanism, applying grease to the lid lock rail for smooth operation
- Test compressor functionality (if applicable) to ensure proper operation and consistent temperature control
- Conduct an 8-12 hour cycle test to validate consistent performance over prolonged use
- Confirm rotor motor stability, check for unusual noise, and assess mechanical integrity during operation
- Inspect the condensation drain and clear any clogs to prevent moisture buildup in the rotor chamber
- Verify RPM accuracy using an NIST-traceable tachometer and perform calibration as necessary
- Check the centrifuge error log to ensure zero recorded issues, confirming system reliability
Technician Inspection and Recertification Checklist: Centrifuges
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task |
Completed? |
Remarks? |
Confirm power supply output voltages match specified requirements |
Completed |
|
Measure and verify voltages at PCB test points for accuracy |
Completed |
|
Confirm rotor compatibility and detection for proper identification |
Completed |
|
Inspect exterior for cosmetic or structural damage |
Completed |
|
Check door mechanism alignment for secure opening and closing |
Completed |
|
Inspect and replace gas strut with OEM part as needed |
Completed |
|
Examine rotor chamber for dents or damage and replace if necessary |
Completed |
|
Replace and confirm functionality of all lid lock switches |
Completed |
|
Lubricate door lock mechanism and lid lock rail for smooth operation |
Completed |
|
Test compressor (if applicable) for proper operation and cooling |
Completed |
|
Run 8-12 hour cycle test to ensure consistent performance |
Completed |
|
Check rotor motor for stability, noise, and integrity |
Completed |
|
Inspect and clear condensation drain to prevent buildup |
Completed |
|
Verify RPM accuracy with NIST traceable tachometer and calibrate |
Completed |
|
Check centrifuge error log for zero recorded issues |
Completed |
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|