Popular Alternative Choice
Thermo Scientific Steri Cycle 370 CO2 Incubator
- Precise CO2 control with high-temperature decontamination.
- In-chamber HEPA air filtration every 60 seconds.
- 140°C on-demand sterilization for easy cleaning.
- Temperature Range: +5°C above ambient to 50°C.
Go to Product Details
Fully Recertified!
Thermo Scientific Napco 8000 CO2 Water Jacketed Incubator
The Thermo Scientific Napco 8000 CO2 Water Jacketed Incubator combines
precise CO2 and temperature control with water-jacketed technology to offer superior stability for cell culture parameters.
The Napco 8000 CO2 Water Jacketed Incubator, with its pinpoint CO2 and temperature control, elevated RH, and use of a thermal conductivity sensor, is an excellent choice for many applications. These incubators feature unsurpassed temperature
stability and superior parameter recovery characteristics, with innovative
continuous contamination control technology.
The Napco 8000's 100% HEPA filtration quickly achieves Class 100 air quality. The HEPA filter entraps
particulate air contaminants and prevents their escape, and filter efficiency
increases as particulates and micro-organisms are entrapped.
Temperature retention and quick recovery are especially important for
researchers working with critical cell cultures. The Thermo Scientific Napco 8000 CO2 Water Jacketed Incubator provides maximum thermal protection and quick recovery from
swings in ambient temperature and power variations.
All of our equipment is fully tested to perform at factory
specifications before being shipped..
The Thermo Scientific Napco 8000 CO2 Water Jacketed Incubator includes:
- > 3 Shelves
- > Drain Tray
- > New HEPA Filter
- > Calibration and Preventative Maintenance Report
- > Power Cord
- > Operator's Manual
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific Napco 8000 CO2 Water Jacketed Incubator Specifications
Temperature:
|
|
Control
|
± 0.1°C
|
Temperature Range
|
Ambient +5°C to 55°C
|
Temperature Uniformity
|
± 0.2°C at 37°C
|
Tracking Alarm
|
User programmable high/low
|
CO2:
|
|
CO2 Control
|
Better than ± 0.1%
|
CO2 Range
|
0-20%
|
Inlet Pressure
|
15 PSIG (1.034 bar)
|
CO2 Sensor
|
T/C
|
Readability
|
0.1%
|
Tracking Alarm
|
User programmable
|
Humidity:
|
|
RH
|
Ambient to 95% at 37°C
|
Humidity Pan
|
0.8 gal (3L) standard
|
Optional Display
|
Displays in 1% increments
|
Inlet Pressure:
|
|
Maximum Per Channel
|
40.8 sq ft.
|
Standard Maximum
|
4,17
|
Temperature Safety:
|
|
Sensor
|
Precision thermistor
|
Controller
|
Independent analog electronic
|
Setability
|
0.1C
|
Construction:
|
|
Water Jacket Volume
|
11.7 gal (43.5L)
|
Interior Volume
|
6.5 cu ft. (184L)
|
Interior
|
Type 304 stainless steel
|
Exterior
|
Cold rolled steel, powder coated
|
Outlet Door Gasket
|
Four-sided, molded magnetic vinyl
|
Interior Door Gasket
|
Feather-edged silicone
|
- Inspect the display for any dim, flickering, or dead pixels, replacing any affected areas to maintain full readability and operational clarity
- Thoroughly examine external regulators for any signs of damage, wear, or malfunction (if applicable), replacing any compromised parts
- Inspect temperature, CO₂, and H₂O sensors (if applicable), ensuring each sensor is functioning correctly and accurately reporting conditions
- Conduct a detailed inspection of all internal wiring for frays, tears, or breakages; replace any compromised wires to ensure safe and reliable connections
- Verify that software settings align with the internal components and calibration specifications, adjusting configurations as needed for accurate operation
- Inspect and clean the fan, ensuring it is free from dust or obstructions; replace if performance is below required standards
- Examine wheels and leveling mounts for structural integrity, ensuring proper balance and stability for safe operation
- Inspect door gaskets and latches for any signs of wear or damage, replacing if necessary to maintain a secure seal and optimal environmental control
- Calibrate temperature accuracy at 37°C using a NIST-traceable probe, confirming it meets required tolerances
- Calibrate CO₂ level to 5% using a Fyrite combustion test kit (if applicable), ensuring precise atmospheric conditions
- Monitor temperature and CO₂ stability over a 24-hour period, verifying consistent operation within specified ranges
- Conduct a thorough cosmetic inspection to confirm all external surfaces and components meet quality standards, free from visible damage or wear
Technician Inspection and Recertification Checklist: Incubator
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Inspect display for dim, flickering, or dead pixels; replace for readability
|
Completed
|
|
Examine external regulators for damage or malfunction; replace compromised parts
|
Completed
|
|
Inspect temperature, CO₂, and H₂O sensors for correct function and accuracy
|
Completed
|
|
Inspect internal wiring for frays, tears, or breakages; replace compromised wires
|
Completed
|
|
Verify software settings align with components and calibration specs; adjust if needed
|
Completed
|
|
Clean fan and remove dust/obstructions; replace if performance is substandard
|
Completed
|
|
Examine wheels and leveling mounts for structural integrity and balance
|
Completed
|
|
Inspect door gaskets and latches for wear or damage; replace to maintain seal
|
Completed
|
|
Calibrate temperature at 37°C with NIST-traceable probe to confirm accuracy
|
Completed
|
|
Calibrate CO₂ level to 5% with Fyrite test kit for atmospheric precision
|
Completed
|
|
Monitor temperature and CO₂ stability over 24-hour period to verify consistency
|
Completed
|
|
Conduct cosmetic inspection to confirm quality of external surfaces and components
|
Completed
|
|
To View More of Our Recertification Protocol Click Here
|
|
|
|
|
|
|