Thermo Scientific NanoDrop Lite Spectrophotometer
The Thermo Scientific NanoDrop Lite Spectrophotometer is a compact, stand-alone UV spectrophotometer designed for basic microvolume measurements of nucleic acids and purified proteins. Featuring a unique sample retention system that uses surface tension to hold 1-2 µL of sample, it eliminates the need for cuvettes. This system, combined with its stainless steel pedestals and optical fibers, ensures accurate and reproducible results. The NanoDrop Lite is equipped with LED technology to measure sample absorption at 260 and 280 nm and corrects for background effects using a reference wavelength at 365 nm. It is capable of determining nucleic acid concentration and purity (via the A260/A280 ratio) and measuring purified protein concentration at 280 nm.
With an absorbance range of 0 to 30 Abs (10 mm equivalent), and a spectral resolution of less than 8.0 nm, it delivers the accuracy expected from NanoDrop instruments. The device is user-friendly, featuring built-in controls and software, an LCD display, and does not require a computer. It also supports connection to an accessory printer for direct label printing. Key specifications include a 3% absorbance accuracy at 1.05 Abs at 260 nm, a minimum sample size of 1 µL, a pathlength of 0.5 mm, and a quick measurement time of under 5 seconds. The NanoDrop Lite, with its small footprint, is an ideal choice for any laboratory benchtop.
All of our equipment is fully tested to perform at factory
specifications before being shipped.
The Thermo Scientific NanoDrop Lite Spectrophotometer includes:
- > CF-1 Calibration Report
- > Power Cord
- > Full 180 Day Parts and Labor Warranty
Thermo Scientific NanoDrop Lite Spectrophotometer Specifications
Absorbance Accuracy
|
3% (at 1.05 Abs at 260 nm)
|
Absorbance Range
|
0 to 30 Abs (10 mm equivalent)
|
Minimum Sample Size
|
1 µL
|
Pathlength
|
0.5 mm
|
Light Source
|
Light emitting diodes
|
Detector Type
|
Silicon photodiode
|
Wavelength Range
|
260 and 280 nm
|
Spectral Resolution
|
< 8.0 nm
|
Detection Limit
|
4 ng/L (dsDNA)
0.12 mg/mL (BSA)
|
Maximum Concentration
|
1,500 ng/L (dsDNA)
45 mg/mL (BSA)
|
Measurement Time
|
< 5 seconds
|
Display
|
LCD
|
Sample Pedestal Material
|
303 stainless steel and quartz fiber
|
Software Compatibility
|
Local (on-board) control only
|
Dimensions
|
6.3 x 4.5 in. (16 x 11.5 cm)
|
Weight
|
1.8 lb. (0.8 kg.)
|
Electrical Requirements
|
6V DC, 18W
|
- Confirm the power supply and PCB output correct, stable voltages across all required connections
- Inspect the display for any dim or dead pixels; replace affected components to maintain full visual clarity and operational accuracy
- Verify full functionality of all buttons, ensuring precise and responsive input control
- Check the XY assembly for proper movement and alignment; thoroughly clean all railings, removing any debris or contaminants
- Confirm stable software communication and connection between the instrument and PC, addressing any connectivity issues
- Inspect and clean or replace all filters as required to ensure optimal airflow and performance
- Examine internal wiring and electrical connections for any breaks, disconnections, or tears; replace or secure any compromised wiring
- Test the lamp’s full functionality, replacing it if any performance issues, dimming, or faults are detected
- Execute the necessary validation protocol, documenting completion with a date on a calibration/verification certificate
- Perform and verify appropriate updates to the PC operating system for compatibility and security
- Conduct a thorough cosmetic inspection, ensuring all external components meet quality and aesthetic standards
Technician Inspection and Recertification Checklist: Instruments
The Following Checklist Is to Be Performed on All Incoming and Outgoing Pieces of Equipment
Task
|
Completed?
|
Remarks?
|
Confirm stable, correct voltage outputs from power supply and PCB
|
Completed
|
|
Inspect display for dim or dead pixels; replace as needed
|
Completed
|
|
Verify functionality of all buttons for precise, responsive control
|
Completed
|
|
Check XY assembly for proper movement; clean railings to remove debris
|
Completed
|
|
Confirm stable software communication between instrument and PC
|
Completed
|
|
Inspect and clean or replace filters to maintain optimal performance
|
Completed
|
|
Examine internal wiring for damage; replace or secure compromised connections
|
Completed
|
|
Test lamp functionality; replace if dimming or faults are detected
|
Completed
|
|
Execute validation protocol and document on calibration/verification certificate
|
Completed
|
|
Update PC operating system for compatibility and security
|
Completed
|
|
Conduct cosmetic inspection to ensure quality of external components
|
Completed
|
|